TE Technology TC-48-20 User Manual
Page 55

55
25. ANALOG OUTPUT MULTIPLIER
Write Command: 2f
Read Command: 63
Interpret:
For writing, multiply desired analog output multiplier by 100
10
then
convert to hexadecimal. For reading, convert to decimal then divide by
100
10
.
Allowable range is 0 to 1.00
10
.
1.00 multiplier is a value of 100
10
0.50 multiplier is a value of 50
10
0.00 multiplier is a value of 0
26. LATCH CLEAR
Write Command: 33
Read Command: NA
Interpret:
Send this command to reset alarm latches.
27. OUTPUT ENABLE
Write Command: 30
Read Command: 64
Interpret:
0 = OFF
1 = ON
28. EEPROM WRITE ENABLE
Write Command: 31
Read Command: 65
Interpret:
0 = OFF
1 = ON
NOTE: this command valid only for firmware versions H and higher. EEPROM WRITE ENABLE is
automatically turned “ON” whenever the controller is turned on. Note: for the TC‐48‐20 only,
regardless of the EEPROM WRITE ENABLE setting, changes made to parameters via keypad are
automatically stored in the EEPROM. This feature can not be turned off.
27. REVISION LEVEL
Write Command: NA
Read Command: 05
Interpret:
08 = H
09 = I
10 = J
11 = K
12 = L
13 = M
14 = N
15 = O
16 = P
17 = Q
18 = R
19 = S
20 = T
21 = U
22 = V
23 = W
24 = X
25 = Y
26 = Z