TE Technology TC-48-20 User Manual
Page 29

29
CONTROL TEMPERATURE box
This shows the set temperature for the controller. The menu options allow for adjustment of the set temperature,
whether a temperature from the secondary thermistor is displayed, and for adjusting the high and low limits of the
set temperature. The values can be changed using the up or down menu buttons or by clicking on the number and
typing in the desired value.
The allowable SET TEMPERATURE is limited to the range defined by the HIGH EXTERNAL SET RANGE and the LOW
EXTERNAL SET RANGE.
The SET TEMPERATURE 2 DISPLAY can be set to OFF to prevent the display of the temperature of the secondary
thermistor, regardless of whether a secondary thermistor is attached to the controller or not. If it is set to
AUTOMATIC, then the controller will automatically display the temperature if a sensor is attached, and turn it off if
no sensor is attached. Setting it to ON will always display the temperature from the secondary thermistor.
The CONTROL SENSOR OFFSET can be used to correct for known sensor errors or differences between the sensed
temperature and the actual temperature.