TE Technology TC-48-20 User Manual
Page 43

43
RS232 Communications Connections
Note: the RS232 cable is not included as a standard accessory. However, TE Technology can provide this as
optional accessory, part number MP‐3023. If using the MP‐3023, connect the leads as follows:
red lead to JP5‐9
clear (or white) lead to JP5‐8
black lead to JP5‐7
Note: other connections to controller have been removed for the sake of clarity.
Note: If you are using a USB to RS232 converter, some converters might induce electronic noise and interfere with
communication. If this occurs, install a 0.001 microfarad, 50V‐ rated, ceramic capacitor across JP5‐7 to JP5‐8.
RS232 Communications Parameters:
A. Baud Rate 115,200
B. No Parity
C. 1 Start Bit 1 Stop Bit
TE
TECHNOLOGY, INC.
TC-48-20
Thermoelectric Temperature Controller
4
3
2
1
12
11
10
9
8
7
6
5
4
3
2
1
TO RS-232 PORT
RECEIVE (into controller)
TRANSMIT (from controller)
GROUND
JP5
JP3
JP4
PC CONNECTOR
PC FUNCTION*
CONTROLLER
9-PIN 25-PIN
PIN
2
3
RECEIVE (RX)
TRANSMIT (TX), JP5-9
PIN 3
2
TRANSMIT (TX)
RECEIVE (RX), JP5-8
PIN 5
7
COMMON
GROUND,
JP5-7
* CHECK YOUR PC MANUAL TO CONFIRM