The SMT384 structures

struct CLKMODE384

bool bExtRef

true = external reference;

false = on-board 10 MHz reference.

The default value is false.

bool bExtClk

true = external clock;

false = on-board VCXO.

The default value is false.

bool bABTrigInvert

true = trigger polarity channel A and B;

false = Non-inverting.

The default value is false.

bool bABTrigExt

true = trigger selection channel A and B;

false = internal trigger.

The default value is false.

bool bCDTrigInvert

true = trigger polarity channel C and D;

false = Non-inverting.

The default value is false.

bool bCDTrigExt

true = trigger selection channel C and D;

false = internal trigger.

The default value is false.

CHSEL ABChSel

Channel A and B selection.

The default value is channelab.

CHSEL CDChSel

Channel C and D selection.

The default value is channelab.

bool bAB2sComplement

True = ADC A and B output 2’s complement samples

False = ADC A and B output binary samples.

The default value is false.

bool bCD2sComplement

True = ADC C and D output 2’s complement samples

False = ADC C and D output binary samples.

The default value is false.

3.3.6

SMT391 interface

The SMT391 interface can be used with a SMT391 plugged in a Sundance carrier board.

This section describes:

the SMT391 functions

Reset

ConfigureADCReg