Test Procedure D
Double Insulated (Class II) Products plus Light-Up test
53. If the FAIL lamp and alarm come on, a flashover has occurred or the leakage current exceeds the
permissible level. Release the Safety and test buttons, then momentarily depress the Flash Fault /
Reset button to reset the fault indications.
Investigate and correct any indicated faults and then fully re-test the product before proceeding.
Light-Up Test
54. Fit appropriate lamps into the product, making sure not to exceed a maximum of 600W loading.
55. Ensure product switches are set to the ON position.
56. Ensure the TEST INHIBIT lamp is not on – if it is, check that the lamp loading is not excessive or disconnect
the product and check for any sources of short circuit in the wiring.
57. Depress the panel mounted SAFETY button and hold it in, then turn the Light-Up Test keyswitch to the ON
position and hold it there. The lamps in the product should light. If not, suspect a faulty lamp. Release the
keyswitch and SAFETY button.
WARNING
ONLY APPLY THE LIGHT-UP TEST ONCE – do not keep flicking the switch on and off as this may cause
nuisance tripping of the supply RCB.
END