7. Flash Fault lamp – lights to show that the product has failed the applied Flash test, an internal audible alarm
will also sound and the test voltage will be removed automatically. This lamp also acts as the Fault Reset
button, which must be momentarily depressed to cancel the fault indication and reset the safety trip circuit.
8. Safety push button – used in conjunction with the Light-Up test to minimize accidental operation.
9. Light-Up test keyswitch – applies mains voltage to the item under test, used as an alternative to the low
voltage continuity/polarity test for products that include ballasts, filter networks, in-line transformers and the
like, and therefore require full mains voltage to check correct operation.
THIS TEST SHOULD ONLY BE APPLIED FOLLOWING SATISFACTORY EARTH AND/OR FLASH TESTS.
10. Test Inhibit lamp – lights to indicate that internal protection circuitry has operated to prevent operation of the
Flash or Light-Up tests. Check that Continuity probes have been removed from lamp holders and/or that total
lamp load does not exceed 600 Watts.