Introducing the CLARE G2000

Descriptions

1. Mains

lead - 240V 13A plug to IEC connector – plug top fuse rated at 3A.


2. Remote Safety Switch – a push-button device used to promote two-handed operation when applying the

high voltage Flash tests. Can be replaced by a Guard Interlock switch when using a safety enclosure.


3. Test Outlet Box – fitted with standard UK and European earthed sockets. An optional ‘Safebloc’ connector is

also available for use with fittings wired with bare ended leads (ie. No plug fitted)


4. Flash Test Probe – a highly insulated safety probe used for applying the Flash test to Double Insulated

(Class II) products. Depressing the red button on the probe handle exposes the test tip.


5. Status

Beacon – used to provide visual indication of instrument status, Green light shows that the instrument

is switched ON and ready for applying tests – Red light shows that a Flash test is in progress.


6. Earth Probe – a high current probe used for applying the Earth tests, the integral green lamp will light to show

a satisfactory earth path. Check the meter reading for actual Earth resistance value.


Accessories 1-6 connect into sockets on the rear panel of the instrument, the appropriate sockets are
correspondingly numbered in FIG.3 on the next page.

7. Continuity Probes – a set of low voltage lamp simulators for checking continuity and polarity of the wiring to

individual lamp-holders. 4 separate probes supplied to suit Bayonet cap (BC) and Edison Screw (ES) fittings
in both standard and small versions (SBC / SES) – other styles can be catered for at extra cost.

.