beautypg.com

000 m – Seaward DO5000 Series User Manual

Page 29

background image

Operating Instructions

Microhmmeter Type DO5000

24

5.6 Measurement with Limits

The measurement limits function permits the fast easy sorting of components.To select the
limits function see section 4.5. Both high and low limit values may be set.If the measured value
is within the set limit boundaries then front panel green PASS LED will illuminate. If the
measured value falls outside the set limits then the red FAIL LED will light, a warning beeper
will also sound and the instrument display will show FAILED with chevron arrows pointing to
the left for value to low and to the right for value to high.

Figure 5.6.1 Measurement display with limits value to low

Figure 5.6.2 Measurement display with limits value to high

AUTO1: 30m

: 10.0A (AVE)

Slow

+ 18.000 m

:

----------------------------------------------------------------------------------------

< Meas Cont

AUTO1: 30m

: 10.0A (AVE)

Slow

+ 18.000 m

:

----------------------------------------------------------------------------------------

>>FAILED>>
Meas Cont