000 m – Seaward DO5000 Series User Manual
Page 29
Operating Instructions
Microhmmeter Type DO5000
24
5.6 Measurement with Limits
The measurement limits function permits the fast easy sorting of components.To select the
limits function see section 4.5. Both high and low limit values may be set.If the measured value
is within the set limit boundaries then front panel green PASS LED will illuminate. If the
measured value falls outside the set limits then the red FAIL LED will light, a warning beeper
will also sound and the instrument display will show FAILED with chevron arrows pointing to
the left for value to low and to the right for value to high.
Figure 5.6.1 Measurement display with limits value to low
Figure 5.6.2 Measurement display with limits value to high
AUTO1: 30m
: 10.0A (AVE)
Slow
+ 18.000 m
:
----------------------------------------------------------------------------------------
<
AUTO1: 30m
: 10.0A (AVE)
Slow
+ 18.000 m
:
----------------------------------------------------------------------------------------
>>FAILED>>
Meas Cont