beautypg.com

Seaward DO5000 Interface User Manual

Page 31

background image

- 31 -

EOT

End of Test. This is asserted when a measurement cycle is
completed.


PASS

If the measurement limits are in use, then this signal will be
asserted with EOT to indicate that the resistance value was
within the limits set.


FAIL LOW/HIGH

If the measurement limits are in use, then this signal will be
asserted with EOT to indicate that the resistance value was
below/above the limits set.


ERROR

This indicates that an error condition occurred during the
previous measurement:


Cause EOT ERROR

Overange

Yes

Yes

Polarity

error

Yes

Yes

Excess

resistance

in Current leads

Yes

Yes

Open Circuit detected

Yes

Yes

Temperature

out

of range

Yes

Yes

Settling

Algorithm

failed to settle

Yes

Yes


Measurement attempted
with empty memory

No

Yes


Measurement attempted
when Datalog is ON, and
maximum No. of readings
has been reached

No

Yes


5.6 Using the PLC Interface

IMPORTANT:
The PLC must be selected as the Remote Interface (See Front Panel Set-up).
When CTRLIN is asserted, the Keyboard is locked.

(If no Remote Interface is selected when CTRLIN is asserted, then the PLC will be
selected automatically. If the RS-232 or IEEE is selected, a warning message will be
displayed, together with an intermittent beep, and the keyboard remains locked until
CTRLIN is released.)

See Figures 1 and 2 for connection details, using the internal supply or an external
supply.

To take a measurement

Make sure that the microhmmeter is displaying the normal measurement screen.
Assert CTRLIN - ‘Rem’ appears on the screen.
Set N0-N3 to select the required set-up memory (See ‘Description of Input/Output
signals’).
Assert MEASURE (Min 10mS) - CTRLOU T is asserted and remains asserted until
CTRLIN is negated. The N0-N3 inputs are read, the appropriate memory is selected,
and the measurement commences. Note that if MEASURE remains asserted,
continuous measurements will occur.
At the end of the measurement, the outputs will be set (See ‘Description of
Input/Output signals’).
Negating CTRLIN will remove ‘Rem’ from the display and unlock the keyboard.