Seaward DO5000 Interface User Manual
Page 19
- 19 -
CALCulate:LIMit:UPPer?
Returns the upper measurement limit in Ohms.
CALCulate:LIMit:STATe
Enables or disables the measurement limit facility.
CALCulate:LIMit:STATe?
Returns “0” (OFF) or “1” (ON).
CALCulate:LIMit:ALARm
Enables or disables the 'failed limit’ audible alarm.
CALCulate:LIMit:ALARm?
Returns “0” (OFF) or “1” (ON).
4.5 Other function Commands
Temperature Compensation Group
SENSe:TCOMpensate:STATe
Enables or disables the temperature compensation facility.
Temperature compensation is not available in FAST mode. Attempting to set the
state “ON” will generate an “execution error”.
SENSe:TCOMpensate:STATe?
Returns “0” (OFF) or “1” (ON).
SENSe:TCOMpensate:MODE
Sets the temperature of the test-piece to either an external Pt100 probe
measurement, or to a fixed, manually-enter ed temperature, the unit of which is
the current temperature unit (ºC or ºF). Range is 0 - 100 ºC (32 - 212°F)
SENSe:TCOMpensate:MODE?
Returns “EXT”, or “MAN”,<manual temperature>.
SENSe:TCOMpensate:REFerence <temperature value>
Sets the reference temperature. This is the temperature to which resistance
measurements are compensated (normally 20ºC). It is entered in the current
temperature unit (ºC or ºF). Range is 0 - 50ºC (32 - 122°F), in integer values of
ºC, or the ºF equivalent.
SENSe:TCOMpensate:REFerence?
Returns the reference temperature in the current temperature unit.
SENSe:TCOMpensate:COEFficient
Sets the temperature coefficient to Copper, Aluminium or to a user-defined value,
which is entered in ppm/ºC (0 - 9999).
SENSe:TCOMpensate:COEFficient?
Returns “CU”,
“USER”,
UNIT:TEMPerature
Sets the temperature unit to ºCelsius or ºFarenheit.
UNIT:TEMPerature?
Returns “CEL” or “FAR”.