beautypg.com

Seaward DO5000 Interface User Manual

Page 19

background image

- 19 -

CALCulate:LIMit:UPPer?

Returns the upper measurement limit in Ohms.


CALCulate:LIMit:STATe or <1|0>

Enables or disables the measurement limit facility.

CALCulate:LIMit:STATe?

Returns “0” (OFF) or “1” (ON).


CALCulate:LIMit:ALARm or <1|0>

Enables or disables the 'failed limit’ audible alarm.

CALCulate:LIMit:ALARm?

Returns “0” (OFF) or “1” (ON).

4.5 Other function Commands

Temperature Compensation Group

SENSe:TCOMpensate:STATe or <1|0>

Enables or disables the temperature compensation facility.

Temperature compensation is not available in FAST mode. Attempting to set the
state “ON” will generate an “execution error”.

SENSe:TCOMpensate:STATe?

Returns “0” (OFF) or “1” (ON).


SENSe:TCOMpensate:MODE or manual temperature
>

Sets the temperature of the test-piece to either an external Pt100 probe
measurement, or to a fixed, manually-enter ed temperature, the unit of which is
the current temperature unit (ºC or ºF). Range is 0 - 100 ºC (32 - 212°F)

SENSe:TCOMpensate:MODE?

Returns “EXT”, or “MAN”,<manual temperature>.


SENSe:TCOMpensate:REFerence <temperature value
>

Sets the reference temperature. This is the temperature to which resistance
measurements are compensated (normally 20ºC). It is entered in the current
temperature unit (ºC or ºF). Range is 0 - 50ºC (32 - 122°F), in integer values of
ºC, or the ºF equivalent.

SENSe:TCOMpensate:REFerence?

Returns the reference temperature in the current temperature unit.


SENSe:TCOMpensate:COEFficient or user coefficient
>

Sets the temperature coefficient to Copper, Aluminium or to a user-defined value,
which is entered in ppm/ºC (0 - 9999).

SENSe:TCOMpensate:COEFficient?

Returns “CU”,, or “AL”,, or
“USER”,. The coefficients are quoted in ppm/ºC.


UNIT:TEMPerature

Sets the temperature unit to ºCelsius or ºFarenheit.

UNIT:TEMPerature?

Returns “CEL” or “FAR”.