Seaward DO5000 Interface User Manual
Page 13
- 13 -
Input Data Format
Numeric data sent to the instrument should be in Ohms or degrees unless stated
otherwise. The value can be in any standard scientific format.
e.g.
3KOHM can be sent as:
3000, 3E3, 3.000E3, 0.3E4 etc.
The units are not sent.
Output Data Format
Non- Measurement queries
ASCII character string
Measurement
queries
IEEE
SDDDD.DDDESDD
RS-232 SDDDD.DDDESDD
Where
S
=
sign
(+/-)
D
=
decimal
digit
(0-9)
E
=
Exponential
Note: In IEEE mode the EOI line is also asserted with the transmission of the
terminator.
Measurement queries always return values expressed in Ohms or Degrees, but the
units are omitted. Resistance values are returned with engineering exponents
reflecting the resolution and units displayed on the LCD screen.
e.g. Display Returned
Value
30.321
Ω
30.321 (30
Ω range)
29.657
K
Ω
29.657E+3
(30K
Ω range)
106.45
m
Ω
106.45E-3
(200m
Ω range)
If a query (?) command results in a “command” or “execution” error, the error value
+9.90E+37 is sent to the output buffer.
Input Message Terminators
In RS-232 mode, all messages sent to the in strument must terminate with either a
terminate a message; the
For IEEE-488 mode, messages can be terminated in one of two ways;
- by sending a
- by asserting the EOI line with the last character sent
It is also permitted to combine these, and assert EOI when sending the
character.
Message termination always forces the command path back to the root-level ready for
the next command message.