Seaward DO5000 Interface User Manual

Page 30

background image

- 30 -

Logic states:

Input - High state

DC or AC voltage with an amplitude between Vmin & Vmax

Input - Low state

no voltage

Output - High state

*closed contact

Output - Low state

*open contact


*Note: these can be reversed by altering the position of links on the PLC card.


25VDC output:

This is provided to drive the PLC inputs/outputs, and is rated at 1.25 ADC. (For the
DO5001 model, it is only available when the instrument is connected to the mains,
and is not supplied by the internal batteries).


Measurement time:

For a basic resistance measurement, the approximate time from the rising edge of the
MEASURE line to the EOT (end of test) signal is:

Slow mode

700 mS

Medium mode

450 mS

Fast mode

240 mS


5.5 Description of Input/Output signals

Inputs:

COM IN

Common line for all inputs.


CTRLIN

Requests remote control of the microhmmeter, and locks
the keyboard. When released, control will revert to the front
panel.


N0-N3

Binary-coded data inputs for selection of set-up memories:


N3 N2 N1 N0

Memory No.

0 0 0 0

0

0 0 0 1

1

0 0 1 0

2

: : : :

:

1 0 0 1

9


MEASURE

Triggers a measurement. If closed momentarily, a single
measurement will result. If left permanently closed,
continuous measurements will result.


Outputs:

The links on the PLC card are factory fitted such that the relay contacts will close
when an output signal is asserted, T hese can be reversed if required.


COM OUT

Common line for all outputs.


ACK

This signal acknowledges the CTRLIN signal, but not until
the first MEASURE input is received. It then remains
asserted as long as CTRLIN is asserted.