2 other information – Seaward Apollo 600 User Manual
Page 29

16.2 Other Information
Parameter
Default
Min
Max
Earth Continuity P/F limit (Ω)
0.1
0.01
19.99
Earth Continuity Duration (s)
5
2
30
Insulation Resistance P/F limit (MΩ)
1
0.01
19.99
Insulation Resistance Duration (s)
5
2
30
Substitute Leakage P/F limit (mA)
0.75
0.25
15
Substitute Leakage Duration (s)
5
2
30
IEC Cord Earth Continuity P/F limit
(ohms)
0.1
0.01
5
IEC Cord Insulation P/F limit (MΩ)
1
0.01
19.99
Touch Current P/F limit (mA)
0.25
0.25
3.50
Touch Current Duration (s)
5
2
30
PE Conductor Current P/F limit (mA)
0.75
0.25
19.99
PE Conductor Current Duration (s)
5
2
30
External Adapter Continuity (Ω)
0.1
0.01
5
External Adapter Leakage (mA)
0.75
0.25
9.99
External Adapter Duration (s)
5
2
30
RCD 30mA trip time (ms)
200
1
2000
RCD 150mA trip time (ms)
40
1
40
Number of tests in a test sequence
1
60
Characters in a test sequence name
1
30
Characters in a User Test name
1
15
Characters in a User Test Unit
15
Total number of Test Sequences
100
Test Results Memory (Mbytes)
300
Test Results (typical)
50000
Sites
100
Characters in Site name
1
15
Locations
100
Characters in Location name
1
15
Users
50
Characters in User name
1
15
Comments List
100
Characters in Comment
25
Characters in Asset ID
1
15