16.2 Other Information

Parameter

Default

Min

Max

Earth Continuity P/F limit (Ω)

0.1

0.01

19.99

Earth Continuity Duration (s)

5

2

30

Insulation Resistance P/F limit (MΩ)

1

0.01

19.99

Insulation Resistance Duration (s)

5

2

30

Substitute Leakage P/F limit (mA)

0.75

0.25

15

Substitute Leakage Duration (s)

5

2

30

IEC Cord Earth Continuity P/F limit
(ohms)

0.1

0.01

5

IEC Cord Insulation P/F limit (MΩ)

1

0.01

19.99

Touch Current P/F limit (mA)

0.25

0.25

3.50

Touch Current Duration (s)

5

2

30

PE Conductor Current P/F limit (mA)

0.75

0.25

19.99

PE Conductor Current Duration (s)

5

2

30

External Adapter Continuity (Ω)

0.1

0.01

5

External Adapter Leakage (mA)

0.75

0.25

9.99

External Adapter Duration (s)

5

2

30

RCD 30mA trip time (ms)

200

1

2000

RCD 150mA trip time (ms)

40

1

40

Number of tests in a test sequence

1

60

Characters in a test sequence name

1

30

Characters in a User Test name

1

15

Characters in a User Test Unit

15

Total number of Test Sequences

100

Test Results Memory (Mbytes)

300

Test Results (typical)

50000

Sites

100

Characters in Site name

1

15

Locations

100

Characters in Location name

1

15

Users

50

Characters in User name

1

15

Comments List

100

Characters in Comment

25

Characters in Asset ID

1

15