Sensor mounted to poorly-grounded or non-conducting surface
7.4
Customer Interface
Cable Requirements
Customer cables that interface to Mercury II Series Encoders must have the following
characteristics:
Twisted pair signals wiring.
Characteristic impedance of 100-120 ohms.
Sufficient wire gauge to meet the minimum voltage requirement at the encoder. For
example: 24AWG gauge wire for a 2m length cable. Recommended cables are 24AWG
gauge wire with 6 twisted pairs.
Single shield cable with a minimum of 90% coverage. Note that a double shielded cable
may be required in high-noise applications.
Power
Supply
5 Volts
O Volts