2 sensor calibration, Sensor calibration – MicroE Mercury II 6000 User Manual
Page 12

Sensor Alignment and Calibration
IM-Mercury_II_6000 Series Rev. 1
Page 11
©2014 MicroE Systems
Mercury
II
6000 Series Encoders
Installation Manual and Reference Guide
5.1.2 Sensor Calibration
Perform calibration when installing the sensor for first time, or if the sensor is subsequently
remounted, or the scale is replaced. This section contains instructions for calibrating the
following:
Linear Scales or Rotary Scales used in applications less than <360°
Rotary Scales used in applications less than >360° without Limit Markers
Confirm Settings
Confirm the following before calibration:
Always perform calibration of the left and right limits while Limit Polarity is set to Limits
Normal mode. See the Left and Right Limits Settings in the Calibration and Align tab of
the SmartPrecision II Software in
Section 5.2.3 Sensor Alignment and Calibration
Select correct Reflective or Non-Reflective Grating Type on Status and Setup tab.