Page 15

background image

4.7.3 Fully Differential

A differential signal has three wires from
the signal source. The signals are Signal
High, Signal Low and Signal Ground
(-IEXC) (which is LLGND).

A differential connection allows you to
connect the CIO-EXP-RTD16 to a signal
source with a ground that is different from
the PC ground, but less than 10V
difference, and still make a true
measurement of the signal.

EXAMPLE: A laboratory instrument with
its own wall plug.

NOTE: There are sometimes voltage
differences in wall GND between outlets.
Measure it to verify that it doesn’t exceed
10V.

Figure 4-7. Differential Voltage Input

11

+I
E

X

C

+S
ENS

E

IE
XC

SENSE

Voltage
Source

Differential Voltage Input

Signal-High

Signal-Low

Ground cable shield at –IEX
terminal only.