Compatible thermocouples

Table 3. Compatible sensor type specifications

Parameter

Conditions

Thermocouple

J: -210 °C to 1200 °C

K: -270 °C to 1372 °C

R: -50 °C to 1768 °C

S: -50 °C to 1768 °C

T: -270 °C to 400 °C

N: -270 °C to 1300 °C

E: -270 °C to 1000 °C

B: 0 °C to 1820 °C

Accuracy

Thermocouple measurement accuracy

Table 4. Thermocouple accuracy specifications, including CJC measurement error. All specifications are (±).

Sensor
Type

Sensor temperature

Accuracy error
maximum (°C)

Accuracy error typical (°C)

Tempco (°C/°C)

J

-210 °C

3.098

1.762

0.040

0 °C

1.282

0.724

1200 °C

1.178

0.684

K

-210 °C

3.318

1.843

0.045

0 °C

1.292

0.730

1372 °C

1.495

0.799

S

-50 °C

1.892

1.058

0.027

250 °C

0.853

0.479

1768 °C

0.734

0.416

R

-50 °C

2.010

1.124

0.025

250 °C

0.844

0.475

1768 °C

0.612

0.347

B

250 °C

2.199

2.192

0.004

700 °C

0.824

0.821

1820 °C

0.471

0.469

E

-200 °C

3.050

1.708

0.038

0 °C

1.465

0.826

1000 °C

1.010

0.564

T

-200 °C

3.226

1.797

0.045

0 °C

1.334

0.754

400 °C

0.856

0.496

N

-200 °C

3.406

1.897

0.035

0 °C

1.300

0.735

1300 °C

0.978

0.571