Troubleshooting
problem
possible cause
remedy
Gel sandwich
leaks while
casting
Dirty or damaged
components
Plates, spacers, and the gasket must be completely clean.
Wash if necessary.
Replace chipped plates (especially if chipped near the spacers).
Check the caster gasket for cuts or cracks and replace if
necessary.
Mis-aligned parts
Check plate and spacer alignment, realign if necessary.
Over-clamping
Turn cam only as far as necessary to create a seal (usually
90 –150°, but up to 180°).
On each spacer apply a light film of Gel Seal compound to
the bottom outside corner only. Do not use silicone grease.
Sample
wells damaged
or irregular
Air bubbles
Remove air bubbles before inserting combs. Slide comb into
solution at an angle. If comb must be removed, add more
monomer solution before reinserting the comb.
Incomplete or
delayed
polymerization
Allow acrylamide gels to set for a minimum of 1 h.
Debris in wells
Rinse out unpolymerized gel with sample buffer.
Comb removal
Remove the comb at a slight angle and very slowly to prevent
damaging the gel.
Agarose gels: Lower the comb no more than 1 cm into the gel.
Incomplete gel
polymerization
Chemicals
Use only recent stocks of the highest-quality reagents.
If the dry ammonium persulfate does not crackle when added
to water, replace with fresh stock.
Increase TEMED or APS concentration, or both.
pH
Solutions with extreme pH values (especially acidic) may not
polymerize.
Oxygen
Remove oxygen from the gel environment: Degas the mono-
mer solution 5–10 min before pouring and then overlay the
gel surface with water-saturated n-butanol.
Temperature
Adjust the gel solution temperature to a minimum of 20 °C,
especially for low %T gels.