beautypg.com

5 resolving the sample – Hoefer SE400 User Manual

Page 28

background image

p20

2.5 Resolving the sample

Electrophoresis parameters for discontinuous
polyacrylamide gels
Gels may be run at either constant current or
constant voltage settings. A constant current
setting is traditionally used with a discontinuous
buffer system so that the rate of electrophoretic
migration remains unchanged throughout the
run. Under constant current conditions, the
voltage increases as the run proceeds. A lower
current setting is recommended for higher reso-
lution. The optimal current level must be deter-
mined empirically; the main factors that must be
balanced are the gel concentration and migration
speed, and the resulting Joule heating and band
distortion. Table 3 lists starting point guidelines
and adjustments for gel thickness, number of
gels, and migration rate.

Table 3. Laemmli buffer system starting point guidelines

Gel thickness*

1.5 mm

Current per gel

25 mA constant current

Starting voltage

80–90 V

Gel length (cm)

model

final voltage (V)

16

SE400

200–250

24

SE410

275–325


* Thicker or thinner gels require proportionally more or less

current. For example, a 0.75 mm gel, which is half as thick
as a 1.5 mm gel, requires half as much current, or 12.5 mA.

The current must be multiplied by the number of gels. For
instance, if a 1 mm 2-gel sandwich is installed, twice as
much current is required than for a single 1 mm gel at the
same voltage.

Note:  The cross section
(and current requirement) is
determined by gel thickness.
The running time is determined
by the length of the plate.

This manual is related to the following products: