INTRODUCTION

Pyramid 8000

®

(8DM) Series compressed breathing

air purifiers are designed to reduce the concentration

of oil, moisture, particulates and carbon monoxide in

the airstream.

When applied at rated capacity and specified operating

conditions as shown in Table III and properly main-

tained, 8DM Series purifiers supply air that meets

OSHA, Canadian Standards Association (CSA) and

Compressed Gas Association (CGA) standards for

compressed breathing air quality, as well as National

Fire Protection Association (NFPA) standards for

medical air quality. Refer to Table 1 for the maximum

allowable carbon monoxide concentration levels at the

purifier inlet.

Air entering the purifier must not be oxygen deficient.

8DM Series purifiers will not increase the oxygen con-

tent of purifier inlet air.

WARNING

To ensure continuing good performance and

safe operation of the purifier, everyone who

installs, operates, maintains or uses it must

read and carefully follow the instructions in

this manual. Installation and maintenance

must be done only by qualified personnel.

SAFETY INSTRUCTIONS

Safety instructions in this manual are bold-faced for

emphasis. The signal words DANGER, WARNING

and CAUTION are used to indicate hazard serious-

ness levels as follows:

DANGER—Immediate hazard which WILL result in

severe injury or death.

WARNING—Hazard or unsafe practice which

COULD result in severe injury or death.

CAUTION—Hazard or unsafe practice which

COULD result in minor injury or in product or prop-

erty damage.

Copies of purifier labels providing important safety in-

formation are included in this manual near correspond-

ing text. Each of the labels is prominently attached to

the purifier before shipment. Contact your distributor

if any label shown in this manual is damaged or miss-

ing from the purifier; identify the label by the part

number printed below it.

Pyramid 8000

®

Series Purifiers (Bulletin 289)

1

Table I

Maximum Allowable Carbon Monoxide Concentration

Purifier Purpose

Maximum

Allowable Inlet

Carbon

Monoxide

Concentration

OSHA

(29 CFR 1910.134 (d)(1))

400 ppm

NFPA 99C* (medical air)

200 ppm

Canadian Standards Association (CSA)*

(CAN3-Z180.01-M85)

100 ppm

Compressed Gas Association (CGA)

(CGA 7.1-1989)

200 ppm

*Concentration of CO and CO

2

combined must not exceed 500 ppm.

8DM Series Instruction Plate

Part No. 44DE127B