5 replaceable parts information, 6 temperature / pressure sensor adjustment – Fluke Biomedical 10100AT User Manual
Page 57
10100AT
Operators Manual
4-4
Use of a different fuse could cause damage to the
instrument.
5. Install the new fuse with the fuse extractor into the instrument by reversing the procedure.
4.5 Replaceable Parts Information
Fluke Biomedical, Radiation Management Services maintains a complete inventory of all normal
replaceable parts (see Table 4-1). To place an order, or to obtain information concerning replaceable
parts, contact the Fluke Biomedical at 440.248.9300.
Table 4-1. Replaceable Parts List
Description Part
Number
20 ft (6.1 m) Coax/Triax Cable
38208
Ion Chamber Stem
32147
Battery Holder
87-3
HVL Filter Kit
37668
Carrying Case, Fitted
37500D
Chamber Holder
37582
HVL Filter Tray
37586
Test Stand Rod
37589
Test Stand Base
37590
Thumbscrew 2715-E-BLACK
Rubber Feet
38471
Bottom Cover Plate
34067
Fuse
19-5
Fuse Extractor
38486
Dust Cap and Chain
878-11-63
Customization Kit with 2 m Cable & Adapters
37594
A.C. Adapter
Call for P/N
4.6 Temperature / Pressure Sensor Adjustment
The measurements from the internal temperature / pressure sensor may be adjusted using the front panel
buttons. Allow the unit’s temperature to equilibrate for at least 1 hour before calibrating the sensor. Verify
that only calibrated thermometers and barometers are used to adjust the 35050AT internal temperature /
pressure sensor. To adjust the sensor measurements:
1. Make sure that the Auto Sense Temperature & Pressure option has been selected.
2. Momentarily depress the Options and Air Density buttons simultaneously.
3. The adjusted/measured temperature and adjusted/measured pressure will be displayed and the
cursor (>) appears next to the value to be adjusted.