Operators Manual
4-2
Figure 4-1. The Model 35050AT Dosimeter Battery Replacement
4.4 Fuse Replacement
1. Twist off the four rubber feet on the bottom of the instrument with finger friction. Remove the
bottom cover plate and set aside.
2. Invert the instrument and tap into the palm of the hand till the battery pack falls out.
3. Look inside the battery pack compartment and find the fuse extractor. It is a loop of plastic tubing
held captive by the fuse body & fuse clips (see Figure 4-2). Grasping the fuse extractor tightly
between your index finger and thumb and as close to the fuse as possible, push away from you
until the fuse pops out. Pull the fuse out of the instrument.
Figure 4-2.
The Model 35050AT Dosimeter Fuse Replacement
4. Replace the fuse with P/N FU-83.
RUBBER FEET
RUBBER FEET
DOSIMETER (REAR VIEW)
BOTTOM COVER PLATE
FUSE
FUSE
EXTRACTOR
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)