Fluke Biomedical 10100AT User Manual
Page 38

10100AT
Operators Manual
2-20
8. During HVL testing, the HVL filters can be positioned on the test stand filter tray.
The kVp and exposure time measurement value
may be affected by the presence of HVL filtering in
the x-ray beam. Be sure to remove the HVL filters
from the “filter tray” when making these
measurements.
NOTE