Electrometer measurement ranges – Fluke Biomedical 10100AT User Manual
Page 10
10100AT
Operators Manual
1-4
Electrometer Measurement Ranges
Table 1-1. Electrometer Measurement Ranges
Ion
Chamber
Units
Effective
Range***
Threshold Resolution
Step Size
Units
Effective
Range***
Threshold Resolution
Step Size
15 cc
R
100
μ to
20
60
μ
1
μ Gy
1
μ to 0.2
.52
μ
0.01
μ
R/s
100
μ to
20
60
μ 1
μ Gy/s
1
μ to 0.2
.52
μ
0.01
μ
R/min
5 m to
1200
3.6 m
50
μ Gy/min
50
μ to
12
31.5
μ
0.5
μ
R/hr
100 m to
72 k
216 m
1 m
Gy/hr
1 m to
720
1.89 m
0.01 m
R/frame**
2
μ to
333 m
1
μ 0.02
μ Gy/frame**
0.02
μ to
3.33m
.008
μ
0.2 n
150 cc
R
10
μ to 2
6
μ 0.1
μ Gy
0.1
μ to
0.02
.052
μ 0.001
μ
R/s
10
μ to 2
6
μ 0.1
μ Gy/s
0.1
μ to
0.02
.052
μ 0.001
μ
R/min
0.5 m to
120
0.36 m
5
μ Gy/min
5
μ to 1.2
3.15
μ 0.05
μ
R/hr
10 m to
7.2 k
21.6 m
0.1 m
Gy/hr
0.1 m to
72
.189 m
0.001 m
R/frame**
0.2
μ to
33 m
0.1
μ 0.002
μ Gy/frame** 0.002
μ
to 0.33 m
0.8 n
0.02 n
150 cc
VLDR
R/s 2
μ to 2*
N/A
0.1
μ Gy/s
0.02
μ to
0.02*
N/A 0.001
μ
R/min
0.1 m to
120*
N/A 5
μ Gy/min
1
μ to
1.2*
N/A 0.05
μ
R/hr
2 m to
7.2 k*
N/A
0.1 m
Gy/hr
0.02 m to
72*
N/A 0.001
m
R/frame**
0.04
μ to
33 m*
N/A 0.002
μ Gy/frame** 0.4
n
0.33 m*
N/A 0.02
n
Electrical
Units
C
1 p to
100 n
0.5 pC
0.01 p
A
1 p to
100 n
250 fA
0.01 p
Values for ion chambers are calculated using nominal sensitivities:
15 cc: 2.4 x 10
8
R/C
150 cc: 2.4 x 10
7
R/C
* Very Low Dose Rate effective range at 5% resolution steps
** At 60 frames/s (1 to 120 frames/ selectable).
*** IEC 61674 effective range at 1% resolution steps