Specifications, Dc volts, Ac volts – Fieldpiece SC77 - True RMS Clamp Meter User Manual
Page 2: Ac current, Dc current, Resistance, Temperature, Continuity, Capacitance, Hi-v(voltage) warning

SPECIFICATIONS
Display:
LCD with a maximum read-
ing of 4000.
Data Hold:
Operating on all ranges
Range(RNG):
Operating on all autor-
anges
Overrange:
"OL" mark indication.
Auto power off:
Approx. 10 minutes.
Measurement rate:
2 times per sec-
ond, nominal.
Operating environment:
32 to 122°F
(0 to 50°C) at <70% R.H.
Storage temperature:
-4 to 140°F
(-20 to 60°C), 0 to 80% R.H. with
battery removed.
Accuracy:
Specifications good in
ambient conditions of 73°F ±9°F
(23°C ±5°) , <75% relative humidity.
Temperature Coefficient:
0.1 ×
(specified accuracy) per °F/°C; 32
to 66
°F (0 to 19°C); 82 to 122°F
(28 to 50°C).
Altitude:
6561.7 Feet (2000 meters).
Jaw opening capability:
38mm con-
ductor.
Power:
Single standard 9-volt battery,
NEDA 1604, JIS 006P, IEC 6F22.
Battery life:
100 hours carbon zinc
200 hours typical with alkaline.
Accessories:
One pair test leads,
one pair of small alligator clips, k-
type thermocouple, 9V battery
(installed), and operating instruc-
tions.
Safety:
Designed to meet
IEC/EN61010-1 and IEC/EN61010-
2-032, CATIII 600V, class II and
pollution degree 2, indoor use
comply with CE. C-Tick.
DC volts
Ranges:
400mV, 4000mV, 40V, 400V,
600V
Resolution:
0.1mV
Accuracy:
±(0.5% rdg + 2 dgt)
Input impedance:
3M
Ω on V inputs,
10M
Ω on mV input
Overload protection:
600VDC or AC
rms
Transient protection:
6kV for 10μ
sec
AC volts
TrueRMS
(50-500Hz)
Ranges:
4V, 40V, 400V, 600V
Resolution:
1mV
Accuracy:
±(2.0% rdg + 8 dgts) 50 ~
500Hz
Input impedance:
3M
Ω
Overload protection:
600V DC or AC
rms
Transient protection:
6kV for 10μ
sec
AC current
TrueRMS
(50-60Hz)
Ranges:
40A, 300A
Resolution:
0.01A
DC Accuracy:
±(3.0% rdg + 10 dgts) 0 ~ 300A /
50 ~ 60Hz
Overload protection:
300AAC
DC current
Ranges:
400μA
Resolution:
0.1μA
Accuracy:
±(1.0% rdg + 2 dgt)
Voltage burden:
1V
Overload protection:
500V DC or AC
rms
Resistance
Ranges:
400
Ω, 4kΩ, 40kΩ, 400kΩ,
4M
Ω, 40MΩ
Resolution:
0.1
Ω
Accuracy:
±(1.2% rdg + 4 dgts) on 400
Ω to
400k
Ω ranges
±(1.5% rdg + 4 dgts) on 4M
Ω range
±(3.0% rdg + 5 dgts) on 40M
Ω
range
Open circuit volts:
-0.45VDC typical,
(-1.2VDC on 400
Ω range)
Overload protection:
500VDC or AC
rms
Temperature
Range:
-30 to 1000°F (-34 to 538°C)
Resolution:
0.1°
Accuracy:
±1°F, 32 to 120°F (0 to 49°C)
±1%+1.5°F, -4 to 750°F (-20 to 399°C)
±2%+4°F, -30 to -4°F (-34 to -20°C)
±2%+4°F, 750 to 1000°F(399 to
538°C)
Sensor type:
K-type thermocouple
Overload protection:
60 VDC or 30
VAC rms
Continuity
Range:
400
Ω
Resolution:
1
Ω
Audible indication:
Less than 25
Ω
Response time:
500ms
Overload Protection:
500VDC or AC
rms
Green LED will be on continuously.
Capacitance
Range:
4μF, 40μF, 400μF, 4mF
1mF = 1000μF
i.e: 1.425mF = 1425μF
Resolution:
1nF
Accuracy:
±(3% rdg + 10 dgts) on
4μF range
±(3% rdg + 5 dgts) on 40μF to
400μF ranges
±(5% rdg + 20 dgts) on 4mF range
Input protection:
500VDC or AC rms
Hi-V(voltage) warning
>30VAC/DC meter beeps, red LED
blinks.
Non-contact voltage
Senses voltage 24V to 600V AC
(50Hz to 60Hz) beeper chirps and big
bright red LED comes on, works when
meter dial is on any range.
Voltage
Resistance
DC Current<4000μADC
Temperature
AC Current
Selecting ranges
For DC voltage, set the meter to the
VDC parameter instead of VAC as
shown above.
For all ranges and functions choose
a range just above the value you
expect. If display reads "OL" (over-
load), select a higher range. If display
shows less than three numbers, select
a lower range for better resolution.
CE
This instrument complies with the
requirements of the following
European Community Directives:
89/336/EEC (Electromagnetic
Compatibility) and 73/23/EEC (Low
Voltage) as amended by 93/68/EEC
(CE Marking).
Capacitance
True RMS
Digital multimeters use two different
types of AC sensing. The most com-
mon is average sensing, normalized to
a true RMS value of a sine wave. The
other is true RMS sensing. The actual
true RMS value is sensed for a wave
form within the limits of the crest factor.
Either sensing method will give the
same results on a clean sine wave but
they may differ on a non-sinusoidal
waveform.
400
1000
µ
Auto
4000m
400m
F C
TEMP
400
F
C
400
1000
µ
Auto
4000m
400m
F C
TEMP
400
F
C
400
1000
µ
Auto
4000m
400m
F C
TEMP
400
F
C
400
1000
µ
Auto
4000m
400m
F C
TEMP
400
F
C
400
1000
µ
Auto
4000m
400m
F C
TEMP
400
F
C
Insure the
temperature
being meas-
ured is stable.
Maintain good
contact
between the
thermocouple
and what's
being meas-
ured.
400
1000
µ
Auto
4000m
400m
F C
TEMP
400
F
C
Disconnect test leads from
voltage before plugging
thermocouple in.
Will display “dsc” if
capacitor needs
discharge.
v16