SPECIFICATIONS

Display:

LCD with a maximum read-

ing of 4000.

Data Hold:

Operating on all ranges

Range(RNG):

Operating on all autor-

anges

Overrange:

"OL" mark indication.

Auto power off:

Approx. 10 minutes.

Measurement rate:

2 times per sec-

ond, nominal.

Operating environment:

32 to 122°F

(0 to 50°C) at <70% R.H.

Storage temperature:

-4 to 140°F

(-20 to 60°C), 0 to 80% R.H. with
battery removed.

Accuracy:

Specifications good in

ambient conditions of 73°F ±9°F
(23°C ±5°) , <75% relative humidity.

Temperature Coefficient:

0.1 ×

(specified accuracy) per °F/°C; 32
to 66

°F (0 to 19°C); 82 to 122°F

(28 to 50°C).

Altitude:

6561.7 Feet (2000 meters).

Jaw opening capability:

38mm con-

ductor.

Power:

Single standard 9-volt battery,

NEDA 1604, JIS 006P, IEC 6F22.

Battery life:

100 hours carbon zinc

200 hours typical with alkaline.

Accessories:

One pair test leads,

one pair of small alligator clips, k-
type thermocouple, 9V battery
(installed), and operating instruc-
tions.

Safety:

Designed to meet

IEC/EN61010-1 and IEC/EN61010-
2-032, CATIII 600V, class II and
pollution degree 2, indoor use
comply with CE. C-Tick.

DC volts

Ranges:

400mV, 4000mV, 40V, 400V,

600V

Resolution:

0.1mV

Accuracy:

±(0.5% rdg + 2 dgt)

Input impedance:

3M

Ω on V inputs,

10M

Ω on mV input

Overload protection:

600VDC or AC

rms

Transient protection:

6kV for 10μ

sec

AC volts

TrueRMS

(50-500Hz)

Ranges:

4V, 40V, 400V, 600V

Resolution:

1mV

Accuracy:

±(2.0% rdg + 8 dgts) 50 ~

500Hz

Input impedance:

3M

Ω

Overload protection:

600V DC or AC

rms

Transient protection:

6kV for 10μ

sec

AC current

TrueRMS

(50-60Hz)

Ranges:

40A, 300A

Resolution:

0.01A

DC Accuracy:

±(3.0% rdg + 10 dgts) 0 ~ 300A /
50 ~ 60Hz

Overload protection:

300AAC

DC current

Ranges:

400μA

Resolution:

0.1μA

Accuracy:

±(1.0% rdg + 2 dgt)

Voltage burden:

1V

Overload protection:

500V DC or AC

rms

Resistance

Ranges:

400

Ω, 4kΩ, 40kΩ, 400kΩ,

4M

Ω, 40MΩ

Resolution:

0.1

Ω

Accuracy:

±(1.2% rdg + 4 dgts) on 400

Ω to

400k

Ω ranges

±(1.5% rdg + 4 dgts) on 4M

Ω range

±(3.0% rdg + 5 dgts) on 40M

Ω

range

Open circuit volts:

-0.45VDC typical,

(-1.2VDC on 400

Ω range)

Overload protection:

500VDC or AC

rms

Temperature

Range:

-30 to 1000°F (-34 to 538°C)

Resolution:

0.1°

Accuracy:
±1°F, 32 to 120°F (0 to 49°C)
±1%+1.5°F, -4 to 750°F (-20 to 399°C)
±2%+4°F, -30 to -4°F (-34 to -20°C)
±2%+4°F, 750 to 1000°F(399 to

538°C)

Sensor type:

K-type thermocouple

Overload protection:

60 VDC or 30

VAC rms

Continuity

Range:

400

Ω

Resolution:

1

Ω

Audible indication:

Less than 25

Ω

Response time:

500ms

Overload Protection:

500VDC or AC

rms

Green LED will be on continuously.

Capacitance

Range:

4μF, 40μF, 400μF, 4mF

1mF = 1000μF
i.e: 1.425mF = 1425μF
Resolution:

1nF

Accuracy:

±(3% rdg + 10 dgts) on

4μF range
±(3% rdg + 5 dgts) on 40μF to
400μF ranges
±(5% rdg + 20 dgts) on 4mF range

Input protection:

500VDC or AC rms

Hi-V(voltage) warning

>30VAC/DC meter beeps, red LED

blinks.

Non-contact voltage

Senses voltage 24V to 600V AC

(50Hz to 60Hz) beeper chirps and big
bright red LED comes on, works when
meter dial is on any range.

Voltage

Resistance

DC Current<4000μADC

Temperature

AC Current

Selecting ranges

For DC voltage, set the meter to the

VDC parameter instead of VAC as
shown above.

For all ranges and functions choose

a range just above the value you
expect. If display reads "OL" (over-
load), select a higher range. If display
shows less than three numbers, select
a lower range for better resolution.

CE

This instrument complies with the

requirements of the following
European Community Directives:
89/336/EEC (Electromagnetic
Compatibility) and 73/23/EEC (Low
Voltage) as amended by 93/68/EEC
(CE Marking).

Capacitance

True RMS

Digital multimeters use two different

types of AC sensing. The most com-
mon is average sensing, normalized to
a true RMS value of a sine wave. The
other is true RMS sensing. The actual
true RMS value is sensed for a wave
form within the limits of the crest factor.
Either sensing method will give the
same results on a clean sine wave but
they may differ on a non-sinusoidal
waveform.

400

1000

µ

Auto

4000m

400m

F C

TEMP

400

F

C

400

1000

µ

Auto

4000m

400m

F C

TEMP

400

F

C

400

1000

µ

Auto

4000m

400m

F C

TEMP

400

F

C

400

1000

µ

Auto

4000m

400m

F C

TEMP

400

F

C

400

1000

µ

Auto

4000m

400m

F C

TEMP

400

F

C

Insure the

temperature
being meas-
ured is stable.
Maintain good
contact
between the
thermocouple
and what's
being meas-
ured.

400

1000

µ

Auto

4000m

400m

F C

TEMP

400

F

C

Disconnect test leads from

voltage before plugging

thermocouple in.

Will display “dsc” if
capacitor needs
discharge.

v16