SPECIFICATIONS

Display: LCD with a maximum read-

ing of 3400.

Analog bar graph: 34 segments with

measurements 20 times per sec-
ond.

Data Hold: Operating on all ranges
Range(RNG): Operating on all autor-

anges

Overrange: "OL" mark indication.
Auto power off: Approx. 10 minutes.
Measurement rate: 2 times per sec-

ond, nominal.

Operating environment: 32 to 122°F

(0 to 50°C) at <70% R.H.

Storage temperature: -4 to 140°F

(-20 to 60°C), 0 to 80% R.H. with
battery removed.

Accuracy: Specifications good in

ambient conditions of 73°F ±9°F
(23°C ±5°) , <75% relative humidity.

Temperature Coefficient: 0.1

×(speci-

fied accuracy) per

°F/°C; 32 to 64°F

(0 to 19°C); 82 to 122

°F

(28 to 50°C).

Altitude: 6561.7 Feet (2000m).
Jaw opening capability: 38mm con-

ductor.

Power: Single standard 9-volt battery,

NEDA 1604, JIS 006P, IEC 6F22.

Battery life: 300 hours typical with

alkaline.

Accessories: One pair test leads,

one pair of small alligator clips, k-
type thermocouple, 9V battery
(installed), and operating instruc-
tions.

Safety: UL, CE, Cat III 600V, IEC

61010-1 (EN61010-1) and
IEC61010-2-032 (EN61010-2-032).

DC volts

Ranges: 340mV, 3400mV, 34V, 340V,

600V

Resolution: 0.1mV
Accuracy: ±(0.5% rdg + 1 dgt)
Input impedance: 3M

Ω on V inputs,

10M

Ω on mV input

Overload protection: 600VDC or AC

rms

Normal mode rejection ratio: >50dB

@ >49Hz

Common mode rejection ratio:

>120dB up to 1000VDC

Transient protection: 6kV for 10μ

sec

AC volts(50Hz - 500Hz)

Ranges: 3.4V, 34V, 340V, 600V
Resolution: 1mV
Accuracy: ±(2.0% rdg + 4 dgts) 50 ~

500Hz

Input impedance: 3M

Ω

Overload protection: 600V DC or AC

rms

Transient protection: 6kV for 10μ

sec

AC current(50Hz - 60Hz)

Ranges: 34A, 300A
Resolution: 0.01A
DC Accuracy:

±(3.0% rdg + 5 dgts) 0 ~ 300A / 50
~ 60Hz

Overload protection: 300AAC

DC current

Ranges: 340μA, 3400μA
Resolution: 0.1μA
Accuracy: ±(1.0% rdg + 1 dgt)
Voltage burden: 1V, (8V on 3400μA

range)

Overload protection: 500VDC or AC

rms

Resistance

Ranges: 340

Ω, 3.4kΩ, 34kΩ, 340kΩ,

3.4M

Ω, 34MΩ

Resolution: 100m

Ω

Accuracy:

±(1.2% rdg + 4 dgts) on 340

Ω to

340k

Ω ranges

±(1.5% rdg + 4 dgts) on 3.4M

Ω

range
±(3.0% rdg + 5 dgts) on 34M

Ω

range

Open circuit volts: -0.45VDC typical,

(-1.2VDC on 340

Ω range)

Overload protection: 500VDC or AC

rms

Temperature

Range: -30 to 1000°F (-35 to 600°C)
Resolution: 0.1°
Accuracy: ±1°F (32 to 120°F); ±0.5°C

(0 to 49°C),

±1% + 1.5°F (-4 to 750°F); ±1% +

1°C (-20 to 400°C),

±2% + 4°F (-30 to -4°F, 750 to

1000°F);
±2% +2°C (-35 to -20, 400 to
540°C)

Sensor type: K-type thermocouple
Overload protection: 60 VDC or 30

VAC rms

Diode test

Test current: Approx. 1.0mA
Resolution: 1mV
Accuracy: ±(2% rdg + 3 dgts)
Open circuit volts: 3.0Vdc typical
Audible indication: Less than 0.2V
Overload protection: 500VDC or AC

rms

Continuity

Audible indication: Less than 300

Ω

Response time: 500ms
Green LED will be on continuously.

Capacitance

Range: 340μF
Resolution: 0.1μF
Accuracy: ±(3% rdg + 5 dgts)
Test frequency: 16Hz
Test voltage: <3.0V
Overload protection: 500VDC or AC

rms

Hi-V(voltage) warning

>30VAC/DC meter beeps, red LED

blinks.

Non-contact voltage

Senses voltage 24V to 600V AC

beeper chirps and big bright red LED
comes on, works when meter dial is on
any range.

Voltage

Resistance

DC Current (μADC)

Temperature

AC Current

Selecting ranges

For DC voltage, set the meter to the

VDC parameter instead of VAC as
shown above.

For all ranges and functions choose

a range just above the value you
expect. If display reads "OL" (over-
load), select a higher range. If display
shows less than three numbers, select
a lower range for better resolution.

Capacitance

CE

This instrument complies with the

requirements of the following
European Community Directives:
89/336/EEC (Electromagnetic
Compatibility) and 73/23/EEC (Low
Voltage) as amended by 93/68/EEC
(CE Marking).

Service

Return any defective SC76 to

Fieldpiece for warranty service along
with proof of purchase. Contact
Fieldpiece for out of warranty repair
charges.

www.fieldpiece.com

Insure the

temperature
being measured
is stable.
Maintain good
contact between
the thermocouple
and what's being
measured.

Disconnect test leads from

voltage before plugging

thermocouple in.