Selecting ranges, Voltage v = ir, Functional specifications – Fieldpiece SC66 - Basic Manual Ranging Clamp Meter User Manual
Page 2: General specifications, Dc current <200µadc, Limited warranty, Obtaining service
selecting ranges
For DC voltage, set the meter to
the VDC parameter instead of VAC
as shown above.
For all ranges and functions
choose a range just above the value
you expect. If display reads “OL”
(overload), select a higher range.
If display shows less than three
numbers, select a lower range for
better resolution.
Voltage
V = ir
volts = current x resistance
Capacitance
resistance
aC Current
temperature
Notes to chart on left:
(1)
50-60Hz
(2)
Input impedance is 560Kohms on
volts ranges, 10Mohm on millivolts
ranges. All ranges will withstand
600VAC/600VDC.
(3)
50Hz-500Hz.
(4)
Protected by PTC (positive tem-
perature coefficient) resistor. If an
overload occurs on microamps DC,
disconnect from circuit for 30 sec-
onds so the PTC can reset.
(5)
Accuracy 1
O
F (32
O
F to 120
O
F);
1%+1.5
O
F (-4
O
F to 750
O
F);
2%+4
O
F (-30
O
F to -4
O
F, 750
O
F to
1000
O
F); all withstand 30V max.
(6)
Test frequency 34Hz at <3.5V. Will
withstand up to 500V.
(7)
Turns on >30V.
(8)
Senses 24VAC to 600VAC.
(9)
0.3VDC typical open circuit voltage.
(3V on 200ohm range). Withstands
up to 500V on any ohms range.
(10)
Turns on <100 ohms within 100mS.
Will withstand up to 500V.
(11)
Holds highest reading.
www.fieldpiece.com
Functional specifications
Accuracy specifications @ 74°F±8°F, relative humidity less than 75%. Accuracy
specified as ±% of reading ± number of least significant digits.
limited warranty
This meter is warranted against
defects in material or workmanship
for one years from date of purchase.
Fieldpiece will replace or repair the
defective unit, at its option, subject to
verification of the defect.
This warranty does not apply to
defects resulting from abuse, neglect,
accident, unauthorized repair,
alteration, or unreasonable use of the
instrument.
Any implied warranties arising
from the sale of a Fieldpiece product,
including but not limited to implied
warranties of merchantability and
fitness for a particular purpose, are
limited to the above. Fieldpiece
shall not be liable for loss of use of
the instrument or other incidental or
consequential damages, expenses,
or economic loss, or for any claim of
such damage, expenses, or economic
loss.
State laws vary. The above limitations
or exclusions may not apply to you.
Obtaining service
Call Fieldpiece Instruments for
one-price-fix-all warranty service
pricing. Send check or money order
for the amount quoted. Send the
meter freight prepaid to Fieldpiece
Instruments. Send proof of date
and location of purchase for in-
warranty service. The meter will be
repaired or replaced, at the option of
Fieldpiece, and returned via least cost
transportation.
General specifications
Indicators: Low battery. Overrange:
“OL” is displayed.
Environmental: Operating tempera-
ture 32°F to 122°F, storage 0°F
to 140°F with batteries removed,
RH<80%.
Measurement rate: 2.5 times/sec
nominal.
Battery: 9V NEDA 1604; 150 hours
typical with carbon zinc (installed).
Auto power off: Meter automati-
cally turns off after approximately
1 hour.
Temperature coefficient: 0.1 X
(specified accuracy) per °F above
82°F and below 64°F.
Industry specifications: IEC/
EN61010-1, IEC/EN61010-2-32,
Cat III (600V for SC66 and SC67)
class 2 and pollution degree 2, in
accordance with IEC-664 indoor
use. UL listed to UL3111. C-Tick
certified.
This instrument complies with
the requirements of the following
European Community Directives:
89/336/EEC (Electromagnetic
Compatibility) and 73/23/EEC (Low
Voltage) as amended by 93/68/EEC
(CE Marking).
Insure the
temperature being
measured is stable.
Maintain good
contact between
the thermocouple
and what’s being
measured.
Disconnect test leads from voltage
before plugging thermocouple in.
DC Current <200µaDC
A PTC resistor, not a fuse, protects
your clamp meter. When an overload
occurs on µADC, do not apply current
for at least 30 seconds to reset the PTC.
CAPACITOR
DISCHARGE
v37
Function Range
Accuracy
/resolution
SC66
SC67
AAC
200
3%+5 / 0.1 (1)
×
×
20
3%+5 / 0.01
VAC(2,3)
200
1.2%+3 / 0.1
600
2%+5 / 1
VDC(2)
20
0.5%+1 / 0.01
200
0.5%+1 / 0.1
µADC(4)
200
1%+1 / 0.1
°F (5)
200°F
1°F / 0.1
1000°F 1%+1.5°F / 1
MFD (6)
200
3%+5 / 0.1
Hi-V (7)
Beep
& LED
NCV (8)
Ohms(9)
200
1%+3 / 0.1
200k
1%+3 / 0.1k
Continuity
(10)
MAX hold
(11)
Diode test
×
×
×
×
×
×
×
×
×
×
×
×
×
×
×
×
×
×
×
×
×
×
×
×
Beep
&LED
×
×
Beep
&LED
+24V