Specifications, Dc volts, Resistance – Fieldpiece SC53 - Dual Display Mini Clamp Meter User Manual
Page 2: Ac current (50hz - 60hz), Dc current (sc52), Capacitance (sc53), Continuity, Diode test (sc52), Temperature (sc53), Temperature

SPECIFICATIONS
Display: 3¾ digit liquid crystal display (LCD) with a maxi-
mum reading of 3999.
Overrange: "OL" or “-OL” mark indication.
Auto power off: 10 minutes.
Operating environment: 32 to 122
°F (0 to 50°C), at <70%
R.H.
Storage temperature: -4 to 140
°F (-20 to 60°C), @ <80%
R.H. with batteries removed.
Accuracy: Specifications good in ambient conditions of
74
°F ±8°F (23°C ±4°C), <75% relative humidity.
Temperature coefficient:0.1
×(specified accuracy) per
°F/°C; 32 to 66°F (0 to 19°C); 82 to 122°F(28 to 50°C).
Altitude: 6561.7 feet (2000m).
Jaw opening capability: 30mm (diameter) conductor
Power: Three lithium-ion CR2032 3V batteries.
Battery life: 100 hours typical for lithium-ion on AAC (top
display), 75 hours typical for lithium-ion on dial functions
(bottom display).
Accessories: One pair vinyl test leads, ATB1 k-type ther-
mocouple (SC53), (3) 3V CR2032 batteries (installed),
ANC4 carrying case, and operating instructions.
Safety: IEC/EN 61010-1 (UL61010-1) CATIII
600V and IEC/EN 61010-2-032 (UL61010-2-032)
CATIII 300V, class II and pollution degree 2
indoor use comply with CE, C-tick certified.
DC Volts
Ranges: 4V, 40V, 400V, 600V
Resolution: 1mV
Accuracy:
±(0.5% rdg + 2 dgts)
Input impedance: 4V: 10M
Ω
40V~600V: 9.1M
Ω
AC Volts (50Hz - 500Hz)
Ranges: 4V, 40V, 400V, 600V
Resolution: 1mV
Accuracy:
±(1.2% rdg + 5 dgts),
±(1.5% rdg + 5 dgts) on 600V range
Input impedance:
4V: 10M
Ω
40V~600V: 9.1M
Ω
Minimum test time for stable reading: 2 seconds
Resistance
Ranges: 400
Ω, 4kΩ, 40kΩ, 400kΩ, 4MΩ, 40MΩ
Resolution: 0.1
Ω
Accuracy:
±(1.0% rdg + 4 dgts) on 400Ω to 400kΩ ranges
±(1.5% rdg + 4 dgts) on 4MΩ range
±(3.0% rdg + 5 dgts) on 40MΩ range
Open circuit volts: -0.45VDC typical, (-1.2VDC on
400
Ω range)
Minimum test time for stable reading: 2 seconds
AC Current (50Hz - 60Hz)
Ranges: 40A, 400A
Resolution: 0.01A
Accuracy:
±(2.0% rdg + 6 dgts)
DC Current (SC52)
Ranges: 400μA, 4000μA
Resolution: 0.1μA
Accuracy:
±(1.0% rdg + 2 dgts)
Voltage burden: 1V (8V on 4000μA range)
Capacitance (SC53)
Range: 4μF, 40μF, 400μF, 4mF
4.000mF = 4000μF
Resolution: 1.0nF
Accuracy:
±(3% rdg + 10 dgts) on 4μF
±(3% rdg + 5 dgts) on 40μF to 400μF
±(5% rdg + 20 dgts) on 4mF (4000μF)
Minimum input range: >100nF
Safety: “dsc” symbol on LCD indicates that the
capacitor must be discharged before testing.
Continuity
Range: 400
Ω
Resolution: 1
Ω
Audible indication: Less than 25
Ω
Response time: 500ms
Diode Test (SC52)
Test current: 1.2mA (approximate)
Accuracy:
±(3.0% rdg + 3 dgts)
Resolution: 10mV
Audible indication: Less than 0.25V
Open circuit volts: 3.0VDC typical
Temperature (SC53)
Range: -30 to 400
°F (-34 to 204°C)
Resolution: 0.1
°
Accuracy: ±1
°F (32 to 120°F); ±0.5°C (0 to 49°C),
±1% + 1.5°F (-4 to 400°F);
±1% + 1°C (-20 to 204°C),
±2% + 4°F (-30 to -4°F);
±2% + 2°C (-34 to -20°C).
Sensor type: K-type thermocouple
Voltage
Amps AC Through Clamp
Resistance
Capacitance
Microamps (for flame diode)
CAT.III
CLAMP
400A
300V
NCV
HOLD
H
H
Disconnect test
leads and slide
TEMP switch to the
right plugging in
thermocouple.
Insure the
temperature being
measured is stable.
Maintain good
contact between the
thermocouple and
what's being
measured.
CAT.
III
CLAM
P
400A
300V
NCV
HOLD
H
H
Temperature
Overload Protection (no fuses to replace)
VAC/DC
600VDC or AC RMS
AAC
400 AAC
Resistance
500VDC or AC RMS
Continuity
500VDC or AC RMS
Diode (SC52)
500VDC or AC RMS
Capacitance (SC53) 500VDC or AC RMS
μADC (SC52)
500VDC or AC RMS
Temperature (SC53) 60VDC or 30VAC RMS
C
US
LISTED