SPECIFICATIONS

Display: 3¾ digit liquid crystal display (LCD) with a maxi-

mum reading of 3999.

Overrange: "OL" or “-OL” mark indication.
Auto power off: 10 minutes.
Operating environment: 32 to 122

°F (0 to 50°C), at <70%

R.H.

Storage temperature: -4 to 140

°F (-20 to 60°C), @ <80%

R.H. with batteries removed.

Accuracy: Specifications good in ambient conditions of

74

°F ±8°F (23°C ±4°C), <75% relative humidity.

Temperature coefficient:0.1

×(specified accuracy) per

°F/°C; 32 to 66°F (0 to 19°C); 82 to 122°F(28 to 50°C).

Altitude: 6561.7 feet (2000m).
Jaw opening capability: 30mm (diameter) conductor
Power: Three lithium-ion CR2032 3V batteries.
Battery life: 100 hours typical for lithium-ion on AAC (top

display), 75 hours typical for lithium-ion on dial functions
(bottom display).

Accessories: One pair vinyl test leads, ATB1 k-type ther-

mocouple (SC53), (3) 3V CR2032 batteries (installed),
ANC4 carrying case, and operating instructions.

Safety: IEC/EN 61010-1 (UL61010-1) CATIII
600V and IEC/EN 61010-2-032 (UL61010-2-032)
CATIII 300V, class II and pollution degree 2

indoor use comply with CE, C-tick certified.

DC Volts

Ranges: 4V, 40V, 400V, 600V
Resolution: 1mV
Accuracy:

±(0.5% rdg + 2 dgts)

Input impedance: 4V: 10M

Ω

40V~600V: 9.1M

Ω

AC Volts (50Hz - 500Hz)

Ranges: 4V, 40V, 400V, 600V
Resolution: 1mV
Accuracy:

±(1.2% rdg + 5 dgts),

±(1.5% rdg + 5 dgts) on 600V range
Input impedance:

4V: 10M

Ω

40V~600V: 9.1M

Ω

Minimum test time for stable reading: 2 seconds

Resistance

Ranges: 400

Ω, 4kΩ, 40kΩ, 400kΩ, 4MΩ, 40MΩ

Resolution: 0.1

Ω

Accuracy:

±(1.0% rdg + 4 dgts) on 400Ω to 400kΩ ranges

±(1.5% rdg + 4 dgts) on 4MΩ range

±(3.0% rdg + 5 dgts) on 40MΩ range

Open circuit volts: -0.45VDC typical, (-1.2VDC on

400

Ω range)

Minimum test time for stable reading: 2 seconds

AC Current (50Hz - 60Hz)

Ranges: 40A, 400A
Resolution: 0.01A
Accuracy:

±(2.0% rdg + 6 dgts)

DC Current (SC52)

Ranges: 400μA, 4000μA
Resolution: 0.1μA
Accuracy:

±(1.0% rdg + 2 dgts)

Voltage burden: 1V (8V on 4000μA range)

Capacitance (SC53)

Range: 4μF, 40μF, 400μF, 4mF
4.000mF = 4000μF
Resolution: 1.0nF
Accuracy:

±(3% rdg + 10 dgts) on 4μF

±(3% rdg + 5 dgts) on 40μF to 400μF

±(5% rdg + 20 dgts) on 4mF (4000μF)
Minimum input range: >100nF
Safety: “dsc” symbol on LCD indicates that the
capacitor must be discharged before testing.

Continuity

Range: 400

Ω

Resolution: 1

Ω

Audible indication: Less than 25

Ω

Response time: 500ms

Diode Test (SC52)

Test current: 1.2mA (approximate)
Accuracy:

±(3.0% rdg + 3 dgts)

Resolution: 10mV
Audible indication: Less than 0.25V
Open circuit volts: 3.0VDC typical

Temperature (SC53)

Range: -30 to 400

°F (-34 to 204°C)

Resolution: 0.1

°

Accuracy: ±1

°F (32 to 120°F); ±0.5°C (0 to 49°C),

±1% + 1.5°F (-4 to 400°F);
±1% + 1°C (-20 to 204°C),
±2% + 4°F (-30 to -4°F);
±2% + 2°C (-34 to -20°C).

Sensor type: K-type thermocouple

Voltage

Amps AC Through Clamp

Resistance

Capacitance

Microamps (for flame diode)

CAT.III

CLAMP

400A

300V

NCV

HOLD

H

H

Disconnect test

leads and slide
TEMP switch to the
right plugging in
thermocouple.

Insure the

temperature being
measured is stable.
Maintain good
contact between the
thermocouple and
what's being
measured.

CAT.

III

CLAM

P

400A

300V

NCV

HOLD

H

H

Temperature

Overload Protection (no fuses to replace)
VAC/DC

600VDC or AC RMS

AAC

400 AAC

Resistance

500VDC or AC RMS

Continuity

500VDC or AC RMS

Diode (SC52)

500VDC or AC RMS

Capacitance (SC53) 500VDC or AC RMS
μADC (SC52)

500VDC or AC RMS

Temperature (SC53) 60VDC or 30VAC RMS

C

US

LISTED