SPECIFICATIONS
Operating environment: 32°F to 122°F; 0°C to
50°C at <75%RH
Storage environment: -4°F to 140°F; -20°C to
60°C at <80%RH with battery removed from
meter.
Battery life: 85 hours typical (alkaline). No
measurable current draw when in "off" position.
Low battery indication: Red LED lights
Battery: 9V
Auto off: Approx. 15 minutes after mode change
Accessories: ATBF1 k-type thermocouple, hand
pump with water trap, 3 C-clips, 5 filters,
2 O-rings, battery (installed),1 screwdriver, oper-
ators manual.
%CO
2
Working range (percent):
0 to 25%
Resolution: 0.1% CO2
Accuracy:
0 to 25% ±0.3% (calculated)
%O
2
Working range (percent):
0 to 25%
Maximum Overload: 30% O2
Resolution: 0.1% O2
Accuracy:
0 to 25% ±0.3%O2 (@72°F (22.2°C) 20.9%O2
ambient)
Stabilization time: under 2 minutes
Temperature
Range (temperature): -58°F to 1000°F; -50°C to
538°C
Resolution: 0.1°F/0.1°C
System Accuracy : ±0.6%+3°F (1.7°C) after field
calibration.
One year limited warranty
This head is warranted to the original purchaser
against defects in material and workmanship for a
period of one year from the date of purchase.
During the warranty period, Fieldpiece will replace
or repair the defective unit, subject to verification of
the defect.
Any damage to the sensor from dirt, mechanical
abuse, or overexposure to damaging chemicals,
including overexposure to carbon monoxide, are
not covered under this warranty. Also not covered
are defects resulting from abuse, neglect, accident,
unauthorized repair, alteration, or unreasonable
use.
ANY IMPLIED WARRANTIES ARISING OUT OF
THE SALE OF A FIELDPIECE INSTRUMENT
PRODUCT, INCLUDING BUT NOT LIMITED TO
IMPLIED WARRANTIES OF MERCHANTABILITY
AND FITNESS FOR A PARTICULAR PURPOSE,
ARE LIMITED TO THE ABOVE. FIELDPIECE
SHALL NOT BE LIABLE FOR LOSS OF USE OF
THE INSTRUMENT OR OTHER INCIDENTAL OR
CONSEQUENTIAL DAMAGES, EXPENSES, OR
ECONOMIC LOSS, OR FOR ANY CLAIM OR
CLAIMS FOR SUCH DAMAGE, EXPENSES, OR
ECONOMIC LOSS.
Local laws vary. Above limitations or exclusions
may not apply to you. This warranty gives you spe-
cific legal rights, and you may also have other rights
which vary by location.
Service
Return any defective AOX2 to Fieldpiece for war-
ranty service along with proof of purchase. Contact
Fieldpiece for out of warranty repair charges.
Fieldpiece Instruments, Inc.
California, U.S.A.
www.fieldpiece.com
K-type High Temperature
Thermocouple
Model: ATBF1
Max: 1000ºF
Description
The ATBF1, K-Type thermocouple can be used
continuously to take temperature readings up to
900ºF and a one time use of 1000ºF. The ATBF1
can be used with any thermometer which accepts a
K-type thermocouple. The ATBF1 also comes with a
wrap tab making it easy to wind and store the ther-
mocouple.
Operation
To use the ATBF1 plug it into any thermometer or
accessory head accepting a K-type thermocouple
and adjust the device to the appropriate settings.
Calibration
Due to variances in the thermocouple wire and
other parts of the system, a field calibration should
be conducted before use. Field calibration typically
gives +/- 1ºF overall accuracy. See Field Calibration
of AOX2 manual.
SPECIFICATIONS
Thermocouple Conductors: K-type Nickel
Chromium/Nickel Aluminum, 2300ºF maximum
(insulation limits max. see Probe insulation).
Accuracy: -50ºF to 545ºF +/- 4ºF, 545ºF to 1000ºF
+/- 0.75%
Range: -50ºF to 900ºF maximum continuos oper-
ation. Single exposure use up to 1000°F.
Probe insulation: While calibration and atmos-
phere will affect maximum useful temperature in
applications, this insulation is designated to
withstand a maximum continuous use at 900°F
(482°C) and a single exposure use up to 1000°F
(583°C).
Plug: K-Type thermocouple male mini plug.
WARNING
The thermocouple may become hot when
testing hot temperatures. Do not handle the
thermocouple when hot.
!
!
v09