Storage

Do not store in areas which contain solvent

vapors. This includes aerosols such as air-freshen-
er, wax polish, window cleaner, and all organic sol-
vents.

Calibration

Because the chemical sensor can be degraded

by exposure to harmful chemicals, dirt and other
contamination, high temperatures, and extremes in
humidity, periodic calibration can maintain accuracy
of the sensor over time. Drift is expected to be less
than 5%/year in most cases but is dependent on
use.

Optional pump

Use the model AOXP2 pump to extract ambient

air samples from hard to reach places or from loca-
tions where the air temperature is high.

For potentially high concentrations of CO, pump

slowly and stop if the measurement approaches
2000PPM.

Carbon monoxide concentrations will begin to be

shown in two or three squeezes. For final reading,
pump until the reading stabilizes--approximately 30
squeezes.

Press MAX on DMM to hold highest reading.

AOXP2

Connect ACM3 in multiple ways

EHDL1 HS series AHDL1 w/ Meter

Cross-sensitivity

The sensor has a permanent unreplaceable filter

built inside the sensor to filter out trace concentra-
tions of SO2, NO2, and most hydrocarbons. If

exposed to high concentrations of harmful chemi-
cals or dirt, the filter can deteriorate and/or impede
diffusion of air to the sensor.

One Year Limited Warranty

This head is warranted to the original purchaser

against defects in material and workmanship for a
period of one year from the date of purchase.
During the warranty period, Fieldpiece will replace
or repair the defective unit, subject to verification of
the defect.

Any damage to the sensor from dirt, mechanical

abuse, or overexposure to damaging chemicals,
including overexposure to carbon monoxide, are
not covered under this warranty. Also not covered
are defects resulting from abuse, neglect, accident,
unauthorized repair, alteration, or unreasonable
use.

ANY IMPLIED WARRANTIES ARISING OUT OF

THE SALE OF A FIELDPIECE INSTRUMENT
PRODUCT, INCLUDING BUT NOT LIMITED TO
IMPLIED WARRANTIES OF MERCHANTABILITY
AND FITNESS FOR A PARTICULAR PURPOSE,
ARE LIMITED TO THE ABOVE. FIELDPIECE
SHALL NOT BE LIABLE FOR LOSS OF USE OF
THE INSTRUMENT OR OTHER INCIDENTAL OR
CONSEQUENTIAL DAMAGES, EXPENSES, OR
ECONOMIC LOSS, OR FOR ANY CLAIM OR
CLAIMS FOR SUCH DAMAGE, EXPENSES, OR
ECONOMIC LOSS.

Local laws vary. Above limitations or exclusions

may not apply to you. This warranty gives you spe-
cific legal rights, and you may also have other rights
which vary by location.

Service

Return any defective ACM3 to Fieldpiece for war-

ranty service along with proof of purchase. Contact
Fieldpiece for out of warranty repair charges.

Fieldpiece Instruments, Inc.

California, U.S.A.

www.fieldpiece.com

Gas

PPM

Cross

sensitivity %

Hydrogen

sulfide

207

1.3

Sulphur

dioxide

208

-0.7

Nitric

oxide

1000

46

Nitrogen

dioxide

20

-4.4

Chlorine

15

-3.8

Hydrogen

200

78

Hydrogen

cyanide

15

-0.5

Hydrogen

chloride

135

-0.7

Removable rubber strip

Use your calibration tool that comes with the

ACM3 to zero adjust.