beautypg.com

Setpoint slope, 7 alarm and counter settings, Alarm settings – Bronkhorst FlowPlot User Manual

Page 17: Alarm and counter settings

background image

BRONKHORST

®

Valve output steps is a special feature and will set the instrument into a mode in which from the last two setpoints

the valve output values will be stored. Sending one of these setpoints to the instrument will immediately result in

setting the valve to the stored output value, than the PID-controller will take over. This option is for very special

applications only. This option is not to be used with pilot-valves because of hysteresis!

Valve overshoot protection will prevent the valve from overshooting, but still open from zero quite fast.

Setpoint slope
Smooth setpoint control can be established using the Setpoint slope. The setpoint will be linearly increased over time

from the old setpoint value to the new setpoint value. A value between 0 and 3000 seconds, with a resolution of 0.1

seconds, can be given to set the time for the integrator on the setpoint signal.

The setpoint will reach its end value after this many seconds:

0.01

* setpoint slope * | new setpoint – old setpoint |

Example: when the setpoint slope is 10 seconds, how long will it take to go from setpoint 20% to 80%?

0.01 * 10 * | 80 – 20 | = 6 seconds

Apart from the normal application purposes, the setpoint slope appeared to be very useful for the internal (software)

controlling mechanism of steering pilot operated valves (e.g. 033, 002AC, 003AC/BC). Due to the mechanical

construction of these valves (containing a small pilot and a bigger main valve), there could be a little overshoot the

first time the main valve opens. Giving setpoint slope a value of about 1 second may reduce this behavior significantly.

This option could be used in combination with special controller feature Automatic slope for pilot valves.

2.6.7

Alarm and counter settings

Alarm settings
Digital instruments contain features to raise an alarm

when a certain event happens. Some of these alarms

indicate an error or warning, showing some

diagnostic information about the state of the

instrument.

On the other hand, the user may select a scenario in

which an alarm is raised. At Mode, the desired

scenario can be selected, which can be one of the

following:

-

Min/max will raise an alarm when the

sensor signal is out of the limits set by

the Minimum and the Maximum value

in percent of the full scale capacity.

-

Reponse will raise an alarm when the

measure of a controller is out of limits

around the setpoint. The limits in

percent of the full scale capacity can be

set at Min and Max.

-

Power-up will raise an alarm when the

instrument is powered-up, e.g. after a

power loss.

To avoid excessive alarms, a Delay time can be set to

delay the alarm event. If the condition for the alarm is

met shorter than this delay time, the alarm is not

raised.

When the alarm is raised, it is possible to

automatically change the setpoint to another value.

The new value can be entered at New setpoint and the automatic setpoint change will be effective when the

checkbox Setpoint change on alarm is checked.

Page 17

FlowPlot

9.17.030