beautypg.com

4 setpoint controller, 5 basic settings, Identification – Bronkhorst FlowPlot User Manual

Page 13: Fluid settings, Output filter, Setpoint controller, Basic settings

background image

BRONKHORST

®

0 – Bus/RS232

Digital setpoint via fieldbus or RS232

1 – Analog input

Setpoint via analog input

3 – Valve fully closed

Valve fully closed and stays closed under all circumstances

8 – Valve fully open

Valve fully opened (purge) and stays opened under all circumstances

11 – Keyboard & FLOW-BUS

Setpoint via E-7000 keyboard, bus or RS232

18 – RS232

Digital setpoint via RS232 neglecting safe state (valve action when no fieldbus

communication present)

For convenience the actual readings of Measure and Valve out of the instrument are shown in percent of maximum

calibrated value and in direct reading.

More detailed information about control modes can be found in document no. 9.17.023.

2.6.4

Setpoint controller

At the Setpoint controller field FlowPlot enables the user to send setpoints to another instrument than the instrument

selected at Channel & Control mode. This can be useful when the connected instrument is not a controller, but a

meter only. It is also possible that this separate controller is connected via another instance of FlowDDE (FlowDDE2 or

vice-versa). The combo box at DDE server will be enabled as soon as FlowPlot determines that there are two FlowDDE

servers running.

Using channel enables the user to connect to a channel of the DDE server. Use the configuration window of the

specific FlowDDE to determine on which channel the instrument can be reached.

At Select setpoint several setpoint steps can be given as a help for taking points for calibration. The setpoints are the

setpoints normally used by Bronkhorst High-Tech to obtain a calibration curve. Eventually use -1% or +1% buttons to

adjust the setpoints.

2.6.5

Basic settings

The Basic tab of the Instrument Settings window is shown in 2.6.1 on page 12.

Identification
The Identification frame is used to display some of the identification properties of the instrument. The Serial number

is a unique number assigned to every instrument. In addition to the serial number, a Service number is assigned to an

instrument which has been serviced by Bronkhorst. Instruments that did not return for service, mostly have this

number set to 000000.

The Model number and the Customer model show the name of the series of the instrument.

The instrument can also be identified by a customizable Usertag, which can be used to easily recognize the

instrument.

Fluid settings
This frame shows which of the eight fluidsets an instrument can contain, is currently active. At Active fluidset the

active fluidset can be easily changed. From the active fluidset, the capacity, the unit, the fluid name and the Controller

speed are shown. The Controller speed slider can be used to adjust the controller speed (it adjusts the controller

gain). The default value is 1.

Output filter
By means of a Dynamic and a Static display factor, the filter for the measured signal going out to either fieldbus or

analog output can be adjusted. The used filter can be influenced by means of a floating point value between 0 and 1.

When the value is 0, the filter is filtering maximal and will not pass any signal change at the input to the output. When

the value is 1, filtering is minimal and the output will be equal to the input signal.

The filter switches to the dynamic factor at large changes of the sensor signal (e.g. sudden pressure or setpoint

changes), while the static factor is used when sensor signal is stable (small signal changes during over time). This

behavior can offer a low-noise sensor signal to the display/output with a relatively low lag in time compared to the

real flow (or other signal). Some advised values are given in the table below.
Page 13

FlowPlot

9.17.030