beautypg.com

AMETEK MX Series Rev: AY User Manual

Page 231

background image

User Manual - Rev AV

AMETEK Programmable Power

MX Series

231

9.6.1 Option –704: MIL-STD-704 rev D & E Tests. (Firmware)

9.6.1.1 General

The MIL704 option is capable of performing most sections of MIL-STD-704 rev D and E in either
of the AC or DC mode.

Note: The Distortion Spectrum tests:


Table I and II, Figure 7
Table III, Figure 12
Table IV, Figure 18

are not supported by the –704 option. These tests require the use of additional
external equipment.

9.6.1.2 Initial Setup

Nominal parameters for the AC Power source are as follows:

Output Voltage

115V L-N

Output Frequency

400 Hz

Nominal parameters for the DC Power source are as follows:

Output Voltage

28V or 270V L-N

A setting outside these nominal values will disable the test and will prevent access to its Menu
screens.

9.6.1.3 Revision

The MIL704 option defaults to Revision E. The System is capable of performing
Revision D test by changing the selection to Revision D. All Revision D tests can be performed
except for the 270V DC TRANSIENT and ABNORMAL. For these tests, Revision E test levels
are used despite the revision D selection.

All levels and timing in this document refers to Revision E. For Revision D, refer to MIL-
STD-704D date 9/30/1980

9.6.1.4 Tests Performed

9.6.1.4.1 STEADY STATE

AC Mode:

1. Steady State Voltage and Frequency test

2. Waveform Distortion test

3. Voltage Modulation test

4. Voltage Unbalance test

5. Phase Unbalance test

6. Frequency Modulation test

7. Transient Voltage low and high test