User Manual – Rev AY

AMETEK Programmable Power

MX Series

123

4.2.13 MEASUREMENT CAL FACTORS Menu

Figure 4-27: MEASUREMENT CAL FACTORS Menu (Series I only)

Figure 4-28: MEASUREMENT CAL FACTORS Menu (Series II only)

The MEASUREMENT CAL FACTORS menu provides access to the measurement calibration
parameters. The parameters apply to the selected mode of operation (AC or DC mode). For
three phase configurations, the PHASE keys toggle between the three calibration screens for
each phase. These parameters are password protected and can only be changed after the
calibration password has been entered. Refer to the calibration section in this manual for details
on performing a calibration.

Note that Series II MX models only have two measurement calibration factors. Series I models
require additional cal factors as shown. The following calibration factors are available from this
menu:

Entry

Description

VOLT FS

Full scale voltage measurement calibration factor.

VOLT 0

Zero offset voltage measurement calibration factor. (Series I
only)

VOLT FS HT

Temperature compensated full-scale voltage measurement
calibration factor. (Series I only)

VOLT 0 HT

Temperature compensated zero offset voltage measurement
calibration factor. (Series I only)

CURR FS

Full scale current measurement calibration factor.

CURR 0

Zero offset current measurement calibration factor. (Series I
only)

CURR FS HT

Temperature compensated full-scale current measurement
calibration factor. (Series I only)

CURR 0 HT

Temperature compensated zero offset current measurement
calibration factor. (Series I only)