6.4.9 Output Calibration Summary
The following Table is a summary of the preceding calibration steps. Note that Series II MX
models have fewer calibration coefficients.
Program the following values in the table and make the adjustments in the OUTPUT
CALIBRATION screen. Select the phase to be calibrated by pressing the PHASE key.
CALIBRATION
PROGRAM VALUES
CALIBRATION VALUE
ADJUST TO
150 VAC range DC Zero
150 VAC range, 0.0 V
VOLT ZERO
0 ± 15 mv DC
150 VAC range Volt FS
120.0 V, 60 Hz
VOLT FS
120
±
0.05 VAC
150 VAC range Volt Hi Freq
120.0 V, 400 Hz
V HI FREQ
120
±
0.05 VAC
Repeat the adjustments at 60 and 400 Hz until the output is within ± 0.05 volts
300 VAC range DC Zero
300 VAC range, 0.0 V,
60 Hz
VOLT ZERO
0 ± 5 mV DC
300 VAC range Volt FS
240.0 V, 60 Hz
VOLT FS
240
±
0.05 VAC
300 VAC range Volt Hi Freq
240.0 V, 400 Hz
VOLT HI FREQ
240
±
0.05 VAC
Repeat the adjustments at 60 and 400 Hz until the output is within
±
0.05 volts
150 VDC range DC Zero
200 VDC range, 0.0 V
VOLT ZERO
0 ± 5 mV DC
150 VDC range Volt +FS
+160.0 V
VOLT FS
+160
±
0.05 VDC
300 VDC range DC Zero
400 VDC range, 0.0 V
VOLT ZERO
0 ± 5 mV DC
300 VDC range Volt
+320.0 V
VOLT FS
+320
±
0.05 VDC
Phase Offset B, C
150 VAC range, 120V,
60 Hz
PHASE OFST
± 0.5
°
Option -413
400 Hz
1800 Hz
R9, R10, R11
INTER HARM FS
20 ± 1.0 Vrms
20 ± 1.0 Vrms
Table 6-3: Output Calibration Table – MX Series I
CALIBRATION
PROGRAM VALUES
CALIBRATION VALUE
ADJUST TO
300 VAC range DC Zero
300 VAC range, 0.0 V,
60 Hz
VOLT ZERO
0 ± 5 mV DC
300 VAC range High Freq
DC Zero, -HF models only.
300 VAC range, 0.0 V,
820 Hz
VOLT ZERO
0 ± 5 mV DC
300 VAC range Volt FS
240.0 V, 60 Hz
VOLT FS
240
±
0.05 VAC
400 VDC range DC Zero
400 VDC range, 0.0 V
VOLT ZERO
0 ± 5 mV DC
Phase Offset B, C
150 VAC range, 120V,
60 Hz
PHASE OFST
± 0.5
°
Option -413
400 Hz, 20VAC
1800 Hz, 20VAC
R9, R10, R11
INTER HARM FS
20 ± 1.0 Vrms
20 ± 1.0 Vrms
Table 6-4: Output Calibration Table – MX Series II