beautypg.com

AMETEK MX Series Rev: AY User Manual

Page 16

background image

User Manual – Rev AY

AMETEK Programmable Power

MX Series

16

Output Parameter

MX30

MX45

MX60

MX90

MX135

V Lo: 300 A
V Hi: 150 A

V Lo: 600 A
V Hi: 300 A

V Lo: 900 A
V Hi: 450 A

Frequency

Range:

Standard:

16 Hz - 819.0 Hz (for –HV option range, 45 Hz – 819.0 Hz)

-LF option:

16 Hz - 500.0 Hz

-HF option:

16 Hz – 905 Hz

Resolution:

0.01 Hz

from 16.00 to 81.91 Hz

0.1 Hz

from 82.0 to 819.0 Hz

Accuracy:

± 0.01 %

±0.25 % for –FC option

Phase (3 phase mode)

Range:

Phase B/C relative to phase A
0.0 to 360.0°

Resolution:

0.1°

Accuracy:

16 Hz - 100 Hz: < 1.5°
100 Hz - 500 Hz: < 2°
> 500 Hz:

< 4°

±0.25 % for –FC option

Ext. Sync Mode (not available with -FC option)

Input:

Isolated TTL input for external frequency control. Requires 5V at 5 mA for
logic high.

Accuracy:

Ext. Sync to phase A with fixed Ext. Sync Frequency input:
16 Hz - 100 Hz: < 2°
100 Hz - 500 Hz: < 3°
> 500 Hz:

< 4°

±0.25 % for –FC option

Programmable Output Impedance (available on MX30-3Pi and MX45-3Pi in three phase mode only)

Range:

R: 1 – 200 mOhm
L: 15 – 200 uH

N/A

N/A

N/A

Resolution:

R: 1 mOhm
L: 1 uH

N/A

N/A

N/A

Accuracy:

10 % FS

N/A

N/A

N/A

Note: Output specifications apply below the Current / Voltage rating lines shown in the V/I

rating chart below.