2.1.4 DC Measurements

Parameter

Range

Accuracy (

±

)

Resolution

Voltage

0 - 400 Volts

0.5 Volts

0.1 Volt

Current

0 - 400 Amps

0.5 Amps

0.01 Amp

Power

0 - 10 kW

30 W

10 W

Note: Accuracy specifications are valid above 100 counts. For current and power measurements,
specifications apply from 2% to 100% of measurement range. Current and Power accuracy and
resolution specifications are times three for MX30-1, MX45-1, MX60, MX90 and MX135 or MX30-3Pi /
MX45-3Pi in single-phase mode.

2.1.5 Harmonic Measurements (Pi controller)

Harmonic measurement specifications apply to MX30-3, MX45-3, MX30-3Pi and MX45-3Pi in three-
phase mode. See notes for other models and configurations.

Parameter

Range

Accuracy (

±

)

Resolution

Frequency fundamental

16.00 - 820 Hz

0.03% + 0.03 Hz

0.01 Hz

Frequency harmonics

MX45-3, MX45-3Pi in 3

phase mode, MX60,

MX90, MX135:

Series I

Series II

32.00 Hz – 6.67 KHz
32.00 Hz – 16 KHz

0.03% + 0.03 Hz

0.01 Hz

MX45-1, MX45-3Pi in 1

phase mode:

Series I

Series II

32.00 Hz - 19.2 KHz
32.00 Hz – 48 KHz

0.03% + 0.03 Hz

0.01 Hz

Phase

0.0 - 360.0°

2° typ.

0.5°

Voltage

Fundamental

0.75V

0.01V

Harmonic 2 - 50

0.75V + 0.3% + 0.3%/kHz

0.01V

Current

Fundamental

0.5A

0.1A

Harmonic 2 - 50

0.15A + 0.3% + 0.3%/kHz

0.1A

Note: For current measurements, specifications apply from 2% to 100% of measurement range.
Current range and accuracy specifications are times three for MX30-1, MX45-1, MX60, MX90 and
MX135 or MX30-3Pi / MX45-3Pi in single-phase mode.

2.1.6 System Specification

Parameter

Specification

External Modulation:

0 to 10%

Synchronization
Input:

Isolated TTL input for external frequency control. Requires 5V at 5 mA for
logic high.

Trigger Input:

External trigger source input. Requires TTL level input signal. Triggers on
negative edge. Response time 80 - 100

µ

s.

Trigger Output:

Programmable through transient list system. 400

µ

s pulse for voltage or

frequency change. Isolated TTL output. Output reverts to Function strobe