3 voltage source programming of ovp – AMETEK DCS-E 1kW Series Programming Manual User Manual

Page 21

background image

DLM-E Series and DCS-E Series

Isolated Analog Programmer Operation

2.3.3 Voltage

Source

Programming of OVP

SW1 Switch Settings


SW1-3 ON (1) = 0–5V
SW1-3 OFF(0) = 0–10V

2

3

6

12

13

1

25

14

16

Set up for voltage source programming of OVP as follows:

1.

Set SW1-3, OVP PROG, to ON (1) for 0-5VDC programming range.

2.

Set SW1-3, OVP PROG, to OFF (0) for 0-10VDC programming range.

3.

Connect the external programming voltage source to the ISOLATED ANALOG
PROGRAMMER connector, J4, with positive to Pin-5 and negative to Pin-12.

4.

Connect Pin-1, ANALOG-CONTROL, of the ISOLATED ANALOG PROGRAMMER
connector, J4, to Pin-2 to enable remote control.

5.

Connect Pin 16, OVP RESISTANCE PROGRAMMING OUTPUT, of the ISOLATED
ANALOG PROGRAMMER connector, J4, to Pin 6 to disable the default OVP
programming connection.

6.

Program the other parameters to the desired limit values.

M51A Option

2-9