Error codes – CIRCUTOR DHB Series User Manual
Page 43

43
7. ERROR CODES
After switching the meter to the network or during the work, messages
about errors can appear.
Messages about errors and their reasons are presented below.
Error message
Description
Overflow of upper value of the measuring range value
or the programmed indication range. The message can
also mean a break in the sensor circuit (thermocouples or
resistance thermometers).
Overflow of lower value of the measuring range value or
the programmed indication range. The message can also
mean a shorting in the sensor circuit (thermocouples or
resistance thermometers).
ErFrt
Communication error with the data memory. Contact the
service workshop.
ErPar
Parameter error. Wrong configuration data.
Manufacturer’s settings will be restored after pressing any
button.
ErdEF
Default settings have been restored. Press any button to
transit to a normal work.
ErFPL
Error of measured values stored by the meter (measured,
maximal and minimal values). Press any button to transit
to a normal work. After pressing the button during 1 sec,
the ErdEF message will be displayed.
Table 12
ErCAo
Lack of calibration of analog outputs. Press any button
to transit to the normal work. Analog outputs will not be
serviced. Contact the service workshop.
ErCAL
Error of calibration. The work is stopped – The meter is
not in the state to carry out measurements in a correct
way. Incorrect checksum of calibration coefficients or
lack of calibration.