Error codes – CIRCUTOR DHB Series User Manual
Page 41
41
7. ERROR CODES
After switching the meter on to the network or during the work, messa-
ges about errors can appear.
Messages about errors and their reasons are presented below.
Table 12
Error message
Description
Overflow of the upper value of the measuring range value
or the programmed indication range.
Overflow of the lower value of the measuring range value
or the programmed indication range.
ErFrt
Communication error with the data memory. One must
contact the service workshop.
ErPar
Parameter error. Wrong configuration data. Manufactu-
rer’s settings will be restored after pressing any push.
ErdEF
Default settings have been restored. One must press any
push to transit to a normal work.
ErFPL
Error of measured values stored by the meter (measured
value, maximal and minimal values). One must press any
push to transit to a normal work. After pressing the push
during 1 sec, the ErdEF message will be displayed.
ErCAo
Lack of calibration of analog outputs. One must press any
push to transit to the normal work. Analog outputs will not
be serviced. One must contact the service workshop.
ErCOd
Erroneous access code to meter parameters. The error
appears in the moment of giving a wrong access code to
meter parameters (only in case when meter parameters
are protected by a password).