CIRCUTOR DHB Series User Manual
Page 17
![background image](https://www.manualsdir.com/files/767765/content/doc017.png)
17
InP 1
Parameter
symbol
Description
Range of changes
tYP1
Kind of the connected input signal
500U – input 500 V.
100U – input 100 V
5A – input 5 A.
1A – input 1 A.
HoUr – current time.
Cnt1
The measurement time is expressed in
seconds. The result on the display pres-
ents the mean value counted in the Cnt1
period. This parameter is not taken into
consideration during the measurement
in the HoUr modes.
1...3600
Table 1
one must hold down the
push-button. Then, the write of the
parameter follows and the display of its symbol alternately with the new
value. The pressure of the
push-button during the change of
the parameter value will cause the resignation of the write.
5.4.2. Changing Floating-point Values
The change is carried out in two stages ( the transition to the next stage
follows after pressing the
push-button:
1) setting values from the range -19999M...99999, similarly as for
integral values;
2) setting decimal point positions (00000., 0000.0, 000.00, 00.000,
0.0000); the
push-button shifts the decimal point to the left,
however the
push shifts the decimal point to the right;
The pressure of the
push-button during the change of the pa-
rameter value will cause the resignation of the write.
5.4.3. Characteristic of Programmed Parameters
Programmed parameters and the range of their quantity changes are
presented in the table below.