Casella CEL CEL-414 User Manual
Page 49

Table 4: Measurement range of CEL-493/3
Range
Freq
Display
Primary
Primary
Pulse
Setting
Wtg
Range
Linearity
Linearity
Range
Range*
Range
Peak
IEC 651
IEC 804
Value
IEC 804
Lo
A
10-111.1 dB
27-91 dB
27-110 dB
86 dB
Linear
10-111.1 dB
45-91 dB
45-110 dB
68 dB
Med
A
40-141.1 dB
43-121 dB
43-140 dB
100 dB
Linear
40-141.1 dB
50-121 dB
50-140 dB
93 dB
Hi
A
50-151.1 dB
53-131 dB
53-150 dB
100 dB
Linear
50-151.1 dB
53-131 dB
53-150 dB
100 dB
Note*: The lower reading is quoted at +10 dB on noise floor.
The upper reading allows crest factor 10.
Table 3: Measurement ranges - All instruments except CEL-493/3)
Range
Freq
Display
Primary
Primary
Primary
Pulse
Setting
Wtg
Range
Linearity
Linearity-
Linearity
Range
Range*
Range*
Range
Peak
IEC 651
IEC 651
IEC 804
Value
(Type 1)
(Type 2)
IEC 804
20
A
20-81,1 dB
27-61 dB
27-67 dB
27-81 dB
57 dB
Linear
20-81.1 dB
45-61 dB
45-67 dB
45-81 dB
39 dB
C
20-81.1 dB
31-61 dB
31-67 dB
31-81 dB
54 dB
50
A
50-111.1 dB
51-91 dB
51-97 dB
51-111 dB
63 dB
Linear
50-111.1 dB
51-91 dB
51-97 dB
51-111 dB
63 dB
C
50-111.1 dB
51-91 dB
51-97 dB
51-111 dB
63 dB
80
A
80-141.1 dB
81-121 dB
81-127 dB
81-141 dB
63 dB
Linear
80-141.1 dB
81-121 dB
81-127 dB
81-141 dB
63 dB
C
80-141.1 dB
81-121 dB
81-127 dB
81-141 dB
63 dB
Note*: The lower reading is quoted at +10 dB on noise floor.
The upper reading allows crest factor 10 for Type 1 instruments, and crest factor 5 for
Type 2 instruments.
General CEL Sound Level Meters Handbook - Page 45