Test Products International 460 User Manual
Page 8
![background image](/manuals/742597/8/background.png)
Input impedance
Approx. 1M Ohm
Sensitivity
50mV to 200V/div 1,2,5 sequence
Accuracy
+/- (3% + 0.05 range/div)
Display modes
CH1, Dual
Maximum Input Voltage
600V
Trigger:
Position
Ver 1; Fixed Position / Ver 3; Adjustable Position
Mode
Free run, normal
Source
Internal (CH1/CH2)
Slope
Positive/Negative
Sensitivity
Real time sampling .... 2 divisions or more
Equivalent sampling .... 3 divisions or more
Trend Plot:
Plot Time
30sec/div to 1-hour/div
Plot Data Type
Max / Min Selectable
Temperature Measurement:
Optional Adapter
A301 (k - Type thermocouple)
Range
-40 F to 500 F
Accuracy
+/- (1.5% + 5 digits)
Pressure / Vacuum Measurement:
Optional Adapter
A620
Range
Pressure 500 psi
Vacuum 30 inHg
Accuracy
Pressure +/- (5% + 5 digits)
Vacuum +/- (1.5% + 5 digits)
Current Measurement:
Optional Adapter 1
A251 (AC AMPS)
Range
0 - 40A / 40 - 400A
Accuracy ( 0 - 40A )
+/- (2.0% + 10 digits , 50 / 60 Hz)
(40 - 400A)
+/- (2.0% + 10 digits , 50 / 60 Hz)
Optional Adapter 2
A256 (AC/DC AMPS)
Range
0 - 40A / 40 - 400A
Accuracy ( 0 - 40A )
+/- (2.0% + 10 digits , 50 / 60 Hz)
(40 - 400A)
+/- (2.0% + 10 digits , 50 / 60 Hz)