beautypg.com

Parr Instrument 6725 User Manual

Page 74

background image

6725

Semi-micro Calorimeter Instruction Manual

14-2

Table 1 – Continued

Auto Preweigh Controls

Automatic Preweigh ID

ON

Automatic Preweigh ID increment

1

Automatic Preweigh ID Number

1

Reporting Controls

Report Width

40

Automatic Reporting

ON

Auto Report Destination

Printer

Individual Printed Reports

OFF

Edit Final Reports

OFF

Recalculate Final Reports

OFF

Use New EE Values
in Recalculation

OFF


Communication Controls

Printer

Printer Type

Parr 1758

Printer Destination

Local (USB)


Balance Port

Balance Port Device

/dev/tty/USB0

Number of Data Bits

8

Parity None
Number of Stop Bits

1

Handshaking None
Baud Rate

9600

Data Characters from Balance

8

Data Precision

4

Transfer Time Out

10




Diagnostics

Calorimetry Parameters

Tolerance (L) Parameters

Preperiod Tolerance (L1)

0.002

Mis-Fire Tolerance (L2)

0.05

Dynamic Postperiod Tolerance (L3) 0.002
Eqlbrm Postperiod Tolerance (L4)

0.002

Preperiod Stability (L5)

2

Preperiod Blackout (L6)

36

Derivative Wghtng Cutoff (L7)

0.02

Preperiod Timeout (L8)

42

Postperiod Timeout (L9)

60

Preperiod Tolerance Jck Calc’d (L10)

0.5


Correction (K) Parameters

K1 0.5
K2 0.00040
K3 1.0
K4 0.0
K5 0.0

Blackout (B) Parameters

Mis-fire Blackout (B2)

6

Derivative Blackout (B3)

0.5

Dynamic Blackout (B4)

6

Equilibrium Blackout (B5)

36

Dynamic Derivative Blackout (B6)

0.02

Dynamic Time Blackout (B7)

20


Data Logger

Data Logger

OFF

Interval in Seconds

10

Data Log Destination

Log File and Printer

Data Log Format

Text