beautypg.com

Parr Instrument 6725 User Manual

Page 41

background image

6725

Semi-micro Calorimeter Instruction Manual

5-14


Automatic Sample ID:
When set to ON the unit will automatically assign sample identification numbers in
accordance with parameters set by the other three keys on this menu. When set to
OFF, the user manually enters each sample ID when prompted to do so.

Auto Sample ID Prefix:
An entry here will be used as a prefix for
all sample IDs, if the Automatic Sample
ID is set to ON. Press this key to access
a sub-menu for entering an alphanumeric
prefix.

Next Auto Sample ID Number:
Establishes the initial sample number for
a series of tests and then shows the next
sample ID which will be assigned. Used
when the Automatic Sample ID is set to
ON. Press this key to access a sub-
menu for entering a numeric increment.

Auto Sample ID Increment:
Establishes the increment between sample numbers; used when the Automatic Sample
ID is set to ON. Press this key to access a sub-menu for entering a numeric increment.

Sample Weight Warning Above:
This key displays and leads to a sub-menu used to set the maximum allowable sample
weight (including spike) in grams. A warning will be given if sample weights above this
value are entered.

Spike Weight Entry Mode:
This key steps through the options for entering spike weights either manually through the
touch screen, network or through the balance port.

Sulfur Entry Mode:
This key toggles steps through the options for entering sulfur correction value either
manually through the touch screen or automatically through the balance port