1 mathematical relations, 11 appendix – Martin Christ Gamma 2-16 LSCplus User Manual
Page 91
![background image](/manuals/739283/91/background.png)
Gamma 1-16 LSC&
Gamma 2-16 LSC&
11 Appendix
Version 04/2013, Rev. 1.1 of 16/12/2013 • sb
91
Translation of the original operating manual
Pos: 230 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/1 10 Anh ang /011 0 Anh ang == == === == == == === == == === == == === == == === == == === == == = @ 6\mo d_1 299 138 116 328 _68. docx @ 35 428 @ 1 @ 1
11 Appendix
Pos: 231 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/1 10 Anh ang /011 0-0 010 M athe matisc he Zusa mm enh äng e @ 6 \mo d_1 299 138 0925 17_ 68. docx @ 35 388 @ 2 @ 1
11.1 Mathematical relations
The automatic processes in the “Programmer module” menu (see chapter
6.5.3 - "Option: freeze-drying with the PGMplus programmer module")
are based on the following considerations:
Fig. 73: Graphical representation of the course of the control set value
Calculation of the control set value and of the gradient for the
temperature:
C/min]
[
sec
sec
sec
sec
tion
current
of
time
tion
tion
previous
of
value
set
tion
current
of
value
set
gradient
C]
[
sec
gradient
time
section
elapsed
tion
previous
of
value
set
value
set
control
Example:
Section
Set values
Section time [h:min]
Temperature [°C]
Preceding
...
30
Current
1:00
60
C/min
0.5
min
60
30
min
60
30
60
C
C
C
gradient
After an elapsed section time of 30 minutes, for example, the control set
value for the temperature is:
C
C/
.
C
value
set
Control
t
45
min
5
0
min
30
30
min)
30
(
Calculation of the control set value for the vacuum:
]
[
10
sec
.)
sec
.
(
10
.
sec
(
10
.)
sec
.
(
10
mbar
value
set
control
time
tion
elapsed
section
current
of
time
section
t
prev
value
set
LOG
t
current
value
set
LOG
t
prev
value
set
LOG
Pos: 232 /01 0 Univ ersal mod ule/ Sei tenwe chsel @ 0\ mod _12 021 1624 431 2_0 .docx @ 1 05 @ @ 1
T/°C
t/h
1000
100
10
1
0.1
0.01
0.001
0.0001
+40
+30
+20
+10
0
-10
-20
-30
Pressure/mbar
Set value of the previous section
Course of the shelf temperature
control set value
Set value of the
current section
Previous
section
Current
section
Following
section