Martin Christ Gamma 2-16 LSCplus User Manual
Page 89
![background image](/manuals/739283/89/background.png)
Gamma 1-16 LSC&
Gamma 2-16 LSC&
10 Technical data
Version 04/2013, Rev. 1.1 of 16/12/2013 • sb
89
Translation of the original operating manual
Connection requirements
(without vacuum pump and
accessories)
Gamma 1-16 LSC
plus
Gamma 2-16 LSC
plus
Electrical connection:
1 x 230 V / 50-60 Hz
(other supply data on request)
1 x 230 V / 50-60 Hz
(other supply data on request)
Protection class:
I
I
IP protection category according to
DIN 60529:
11
11
Nominal power:
on request
1.45 kVA
Nominal current:
on request
6.5 A
Mains fuse:
10 A
12 A
Vacuum pump connection:
230 V, 50/60Hz, 2 A max.
230 V, 50/60 Hz, 2 A max.
Filling quantity:
Gamma 1-16 LSC
plus
Gamma 2-16 LSC
plus
Refrigerant
See name plate
See name plate
Physical data (without vacuum
pump and accessories)
Gamma 1-16 LSC
plus
Gamma 2-16 LSC
plus
Dimensions
- height:
- width:
- depth:
485 mm
860 mm
650 mm
(incl. vacuum flange connection)
485 mm
860 mm
650 mm
(incl. vacuum flange connection)
Weight:
approx. 135 kg
approx. 160 kg
Noise level according to DIN
45635:
56 dB(A)
56 dB(A)
EMC according to EN 55011:
Class B
Class B
Heat emission:
0.75 kW min.
1.3 kW max.
1.6 kW min.
2.2 kW max.
Equipment connections
Gamma 1-16 LSC
plus
Gamma 2-16 LSC
plus
Vacuum connection:
Small flange connection DN25KF
(ISO 28403, DIN 2861)
Small flange connection DN25KF
(ISO 28403, DIN 2861)
Condensate drain valve:
Hose nozzle DN10
(outside diameter 12 mm)
Hose nozzle DN10
(outside diameter 12 mm)
Aeration valve:
Hose nozzle DN6
(outside diameter 10 mm max.)
Hose nozzle DN6
(outside diameter 10 mm max.)
Vacuum sensor:
SUB D-9
VSP 62
SUB D-9
VSP 62
Data interface (LAN):
RJ 45
RJ 45