7 Malfunctions and error correction
78
Version 04/2013, Rev. 1.1 of 16/12/2013 • sb
Translation of the original operating manual
Pos: 178 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 010 -00 30 Eisko nde nsat orte mp erat ur Alp ha bis Delta @ 6\ mod _12 991 375 173 10_ 68.d ocx @ 351 76 @ 3 @ 1
7.1.3 Insufficient ice condenser temperature
CAUTION
Ensure sufficient ventilation. Do not place any paper, cloth, or similar
material behind or under the unit, since otherwise the air circulation will be
impaired.
The refrigeration unit is equipped with a protective device against
overpressure in the refrigeration system and with a thermal motor
protection switch.
The protective devices trip
• when the ambient temperature is too high
• when the air circulation of the heat exchanger of the refrigeration
system is insufficient
• when the refrigeration system is overloaded.
In these cases, the refrigeration unit will be switched off automatically. If the
permissible operating conditions are re-established after a cool-down
phase of several minutes, the refrigeration unit will be switched on again
automatically.
The malfunctions are displayed in the process and equipment information
window.
The minimum ice condenser temperature of approx. -55°C or approx. –
85°C (depending on the type of freeze-dryer) is reached when the ice
condenser is not loaded and the ice condenser chamber is evacuated.
Pos: 179 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 180 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 181 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 020 Proz ess un d F ehle rm eldun gen @ 6\ mod _12 991 375 176 38_ 68.d ocx @ 351 84 @ 2 @ 1
7.2 Process and error messages
The LSC
plus
control system displays the complete process and error
messages, which is why they are not included in this operating manual.
You can order these documents on the Internet at www.martinchrist.de by
way of the form "Service request".