3 option: freeze-drying with the pgm, Programmer module – Martin Christ Gamma 2-16 LSCplus User Manual
Page 61
![background image](/manuals/739283/61/background.png)
Gamma 1-16 LSC&
Gamma 2-16 LSC&
6 Operation
Version 04/2013, Rev. 1.1 of 16/12/2013 • sb
61
Translation of the original operating manual
condensate water or another medium, the opening time of the drain
valve can be pre-selected in this dialog box.
• Leakage test (4)
This function is optional (see chapter 6.6 - "Optional extensions"). This
button displays the parameters that are used for the leakage test. In
addition, the test time and the chamber volume must be stated
(depending on the accessories that are used).
Numerical values can be edited with the aid of a numerical keypad:
1 Set value display
2 Button for changing the
sign (e.g. when entering
temperature values)
3 " " (infinite)
4 Button for editing the
value in pre-defined
steps
5 Selection of a possible
maximum or minimum
value
Input of a value via the
numerical keypad
7 Button for deleting the
displayed value
Fig. 55: Editing set values
• Confirm the new value and quit the numerical keypad by pressing the
button “Accept”.
• Confirm the input and quit the dialog box via the button "Accept".
• If the dialog box is closed by the button "return", the changes will be
discarded.
Pos: 139 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 140 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 60 Bet rieb/ LSCplus/ 060 -00 50- 003 0 Ge frie rtroc knu ng mi t Pro gra mmg ebe r PGM plus Lab or @ 6\m od_ 129 914 275 4483 _68 .docx @ 3 671 1 @ 3 @ 1
6.5.3 Option: freeze-drying with the PGM
plus
programmer module
Unlike in the manual mode, an entire freeze-drying process can be
executed fully automatically and under reproducible conditions with the aid
of the PGM
plus
programmer module.
NOTE
The PGM
plus
programmer module is an option that is not available as
standard. If the programmer module is not enabled, programs can be
created and edited for demonstration purposes only.The execution of a
program, however, is not possible in this case!
Freeze-drying programs are divided into sections (1). Every section in the
program has certain set values (2). A program must include a minimum of
two sections, and the maximum number of possible sections is 64.
32 program storage locations are available.
In every program, the system always displays four consecutive sections in
order to show their connection.