Labconco WaterPr BT Systems 90151 Series User Manual
Page 26

18
How to Calibrate the Flowrate from the WaterPro® BT System (F02)
Before calibrating the Product flowrate from the WaterPro® BT System, you will need a 1 Liter graduated cylinder
to measure the total volume of water that will be dispensed. The Final Filter should be installed.
1
Press the Main and “-“ Buttons together
to enter the menu. The Display will show
F01.
2
Press the Main Button once. The Display
will show F02 and the 60 second timer.
The graduated cylinder will be blinking.
3
Press the Dispensing Button once. The
system will dispense water for 60 seconds.
The Display will show the timer counting
down.
NOTE: If you press the Dispensing
Button again or if the Tank Level
display is at the 10% level before 60
seconds have elapsed, then the system
will stop dispensing water. Calibration
of the flowrate is cancelled.
Continued on next page